Performance Verification Procedures
SDA 601 Serial Digital Analyzer
3- 5
2.
Switch the equipment on and allow a 20-minute warm-up. Set the front panel
controls of the Tektronix 1780 (or 1781) as listed in Table 3--2.
Table 3- 2: Initial Settings for the Video Measurement Set
Control
Setting
LEFT DISPLAY
PIX
RIGHT DISPLAY
WFM
INPUT
CH A
REF
INT
FILTER
FLAT
WFM HORIZONTAL
TWO, LINE
GAIN
NORMAL
MAGNIFIER
OFF
The picture on the Left Display of the Video Measurement Set should
resemble Figure 3--2. If the center portion of the display does not contain the
darker OSD (on-screen display) rectangle, press the SDA601
Insert On/OFF
button to toggle the OSD On.
Figure 3- 2: AUX Output Containing the On-Screen Display (OSD)
3.
To enter the SDA601 Utility menu, hold the
Lock Out
key down and press
the
ON
button. Release the
ON
button, but continue to press the
Lock Out
key
until the display resembles Figure 3--3.
Содержание SDA601
Страница 4: ......
Страница 10: ......
Страница 15: ...Specifications...
Страница 16: ......
Страница 21: ...Operating Information...
Страница 22: ......
Страница 29: ...Operating Information SDA 601 Serial Digital Analyzer 2 7 Figure 2 5 A Printed Analysis Report...
Страница 49: ...Performance Verification...
Страница 50: ......
Страница 69: ...Adjustment Procedures...
Страница 70: ......
Страница 78: ...Adjustment Procedures 4 8 SDA 601 Serial Digital Analyzer...
Страница 79: ...Maintenance...
Страница 80: ......
Страница 93: ...Replaceable Electrical Parts...
Страница 94: ......
Страница 104: ......
Страница 105: ...Diagrams...
Страница 106: ......
Страница 108: ...Diagrams and Circuit Board Illustrations 7 2 SDA 601 Serial Digital Analyzer Data Path Diagram...
Страница 111: ......
Страница 122: ......
Страница 123: ...Replaceable Mechanical Parts...
Страница 124: ......
Страница 129: ...Index...
Страница 130: ......