Operation
Control acquisitions
When two or more displays are open, the selected display controls the acquisition system. The acquisition requirements for
the selected display set the acquisition parameters. While the acquisition parameters are optimized for the selected display,
they might not be ideal for other displays. In this case, the other displays might be unable to maintain optimum performance
and consequently might be unable to produce results. (This also happens when you analyze recalled data: measurements
attempt to deliver results, but may not be able to comply with all their settings.)
You can specify which display controls the acquisition parameters by clicking anywhere in that display. When there are
multiple displays, the highlighted display controls the acquisition parameters.
In this example, there are two displays, a
Spectrum display and a DPX Spectrum
display. The highlighted title bar on the
Spectrum display indicates that it is the
selected display.
The check mark indicator in the upper left
corner of the display indicates that acquisition
is optimized for the Spectrum display.
Some displays do not control the acquisition. These displays just show the captured data. Some displays share the same
parameters. If any of one of these displays in such a family controls the acquisition, then the acquisition parameters are
optimized for all of the displays in that family.
70
RSA7100 Series Quick Start User Manual
Содержание RSA7100 Series
Страница 1: ...x RSA7100 Series Real Time Spectrum Analyzer ZZZ Quick Start User Manual 071 3504 10 ...
Страница 2: ......
Страница 6: ......
Страница 20: ...Preface xiv RSA7100 Series Quick Start User Manual ...
Страница 102: ...Operation 82 RSA7100 Series Quick Start User Manual ...