Adjustment Procedures
RSA3408B Analyzer Service Manual
4- 17
To run the gain calibration, do the following:
NOTE
. When you run the gain calibration during signal acquisition, the
calibration begins after the acquisition is completed.
1.
Press the
Cal
key on the front panel (see Figure 4--7).
2.
Press the
Calibrate Gain
side key.
The calibration runs. It takes several seconds to complete the process.
When you want to perform the gain, center offset, and DC offset calibration
together, press the
Calibrate All
side key.
3.
If you press the
Auto Calibration
side key to select
Yes
, the calibration will
run automatically any time the analyzer gain drifts to an uncal state.
Cal menu
Cal key
Calibrate all
Gain calibration
Center offset calibration
DC offset calibration
Auto Calibration
(Select Yes or No)
Figure 4- 7: Calibration menu
Содержание RSA3408
Страница 4: ......
Страница 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Страница 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Страница 20: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Страница 26: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Страница 82: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Страница 163: ...Removal and Installation Procedures RSA3408B Analyzer Service Manual 5 45 Figure 5 23 Motherboard removal...
Страница 200: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Страница 202: ...6 4 RSA3408B Analyzer Service Manual...
Страница 204: ...6 6 RSA3408B Analyzer Service Manual...
Страница 210: ......
Страница 211: ......
Страница 215: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 11 1 2 RF4 RF2 RF3 RF6 3 4 5 RF5 6 Figure 7 4 RF modules...
Страница 223: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 19 1 2 3 4 5 6 7 8 9 10 11 Figure 7 8 Disk modules Option 06 2...
Страница 231: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 27 1 2 3 4 5 6 7 8 9 10 11 12 Figure 7 12 Chassis modules 2...
Страница 235: ......
Страница 236: ......