Performance Tests
3- 12
RSA3408B Analyzer Service Manual
Auto Tests
The auto tests use internal routines to verify specifications for the following
characteristics:
H
Baseband flatness
H
IF flatness
H
RF flatness
H
Spurious response
Use the service PV menu and procedure described below to set parameters and
perform the tests.
Selects the PV (performance verification) test items and bands. Access this menu
by pressing
Cal
>
Service...
>
Password (270833)
>
PV
(see Figure 3--6).
Figure 3- 6: Auto test initial screen
The PV menu has the following controls:
PV Select.
Selects the test items as shown in Table 3--4.
Service PV Menu
Содержание RSA3408
Страница 4: ......
Страница 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Страница 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Страница 20: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Страница 26: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Страница 82: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Страница 163: ...Removal and Installation Procedures RSA3408B Analyzer Service Manual 5 45 Figure 5 23 Motherboard removal...
Страница 200: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Страница 202: ...6 4 RSA3408B Analyzer Service Manual...
Страница 204: ...6 6 RSA3408B Analyzer Service Manual...
Страница 210: ......
Страница 211: ......
Страница 215: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 11 1 2 RF4 RF2 RF3 RF6 3 4 5 RF5 6 Figure 7 4 RF modules...
Страница 223: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 19 1 2 3 4 5 6 7 8 9 10 11 Figure 7 8 Disk modules Option 06 2...
Страница 231: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 27 1 2 3 4 5 6 7 8 9 10 11 12 Figure 7 12 Chassis modules 2...
Страница 235: ......
Страница 236: ......