Adjustment Procedures
RSA3408B Analyzer Service Manual
4- 19
The DC offset calibration cancels DC offset that appears at 0 Hz in the baseband.
When you change the amplitude setting and the DC offset is too obvious, run the
DC offset calibration.
When DC offset appears at 0 Hz in
the baseband, run the calibration.
Figure 4- 9: DC offset
1.
Press the
Cal
key on the front panel (see Figure 4--7).
2.
Press the
Calibrate DC Offset
side key.
The calibration runs. It takes several seconds to complete the process.
When you want to perform the gain, center offset, and DC offset calibration
together, press the
Calibrate All
side key.
DC Offset Calibration
Содержание RSA3408
Страница 4: ......
Страница 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Страница 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Страница 20: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Страница 26: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Страница 82: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Страница 163: ...Removal and Installation Procedures RSA3408B Analyzer Service Manual 5 45 Figure 5 23 Motherboard removal...
Страница 200: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Страница 202: ...6 4 RSA3408B Analyzer Service Manual...
Страница 204: ...6 6 RSA3408B Analyzer Service Manual...
Страница 210: ......
Страница 211: ......
Страница 215: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 11 1 2 RF4 RF2 RF3 RF6 3 4 5 RF5 6 Figure 7 4 RF modules...
Страница 223: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 19 1 2 3 4 5 6 7 8 9 10 11 Figure 7 8 Disk modules Option 06 2...
Страница 231: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 27 1 2 3 4 5 6 7 8 9 10 11 12 Figure 7 12 Chassis modules 2...
Страница 235: ......
Страница 236: ......