Adjustment Procedures
4- 24
RSA3408B Analyzer Service Manual
2.
Enter the calibration menu:
a.
Press
Service...
>
Password
in the Cal menu.
b.
Enter
270833
and then press
Enter
using the numeric keypad.
c.
Press the
Cal
side key.
d.
Set the GPIB address of the generator:
H
Press the
Config...
side key.
H
Press the
SG Type
side key to select the manufacturer of the
generator for the GPIB command system: HP or Anritsu.
H
Press the
SG Address
side key and set the generator address.
H
Press the
Cancel--Back
side key.
3.
Perform the 10 MHz reference calibration:
Press the
10MHz Reference
side key to run the calibration.
4.
Disconnect the test equipment:
a.
Disconnect the cable from the analyzer input.
b.
Turn off the analyzer and generator.
c.
Disconnect the GPIB cable.
Содержание RSA3408
Страница 4: ......
Страница 14: ...Service Safety Summary x RSA3408B Analyzer Service Manual...
Страница 18: ...Introduction xiv RSA3408B Analyzer Service Manual...
Страница 20: ...Specifications 1 2 RSA3408B Analyzer Service Manual...
Страница 26: ...Theory of Operation 2 6 RSA3408B Analyzer Service Manual...
Страница 82: ...Performance Tests 3 56 RSA3408B Analyzer Service Manual...
Страница 163: ...Removal and Installation Procedures RSA3408B Analyzer Service Manual 5 45 Figure 5 23 Motherboard removal...
Страница 200: ...Diagrams 6 2 RSA3408B Analyzer Service Manual...
Страница 202: ...6 4 RSA3408B Analyzer Service Manual...
Страница 204: ...6 6 RSA3408B Analyzer Service Manual...
Страница 210: ......
Страница 211: ......
Страница 215: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 11 1 2 RF4 RF2 RF3 RF6 3 4 5 RF5 6 Figure 7 4 RF modules...
Страница 223: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 19 1 2 3 4 5 6 7 8 9 10 11 Figure 7 8 Disk modules Option 06 2...
Страница 231: ...Replaceable Parts RSA3408B Analyzer Service Manual 7 27 1 2 3 4 5 6 7 8 9 10 11 12 Figure 7 12 Chassis modules 2...
Страница 235: ......
Страница 236: ......