Speci
fi
cations
Table 3: Spectral analysis
Characteristic
Description
Measurement
Frequency
Sets the measurement frequency of the display. The measurement frequency can be set by either
entering a speci
fi
c frequency or by selecting a channel number.
Units
MegaHertz (MHz) or Channel Number
Resolution
1 kHz 1 channel
Range
30 MHz to 2500 MHz
Frequency Span
Sets the frequency range covered by the spectral display.
Units
kiloHertz (kHz), MegaHertz (MHz)
Resolution
1 kHz
Range
10 kHz to 2470 MHz, continuously variable or 1-2-5 steps
Resolution Bandwidth
(RBW)
Width of the resolution bandwidth
fi
lter
Units
Hertz (Hz), kiloHertz (kHz), MegaHertz (MHz)
Resolution
Uses 1-3 steps to cover the range
Range
10 Hz to 6 MHz
When in Auto mode, the instrument automatically sets the RBW. The RBW is set as a combination
of 1-3 steps at the 1, 2, 5 span settings, and a linear range between 1, 2, 5 span settings, for
a nominal value of approximately span/125.
76
RF Scout Quick Start User Manual