Performance Veri
fi
cation
6.
Set the probe to the 10 mA/mV range.
7.
Set the oscilloscope to:
Probe setup to current, 10X attenuation
Vertical sensitivity to 500 mA/div
Horizontal to 1 ms/div
8.
Repeat steps 2 through 5 for the 10 mA/mV
range.
9.
Disable the generator output.
End of procedures.
47
P6021A Current Probe Instruction Manual