Performance Veri
fi
cation
Equipment Setup
Use the following procedure to set up and warm up the equipment to test the probe.
1.
Turn on the oscilloscope.
2.
Connect the probe to any channel of the
oscilloscope.
3.
Set the oscilloscope coupling to AC.
4.
Power on the remaining test equipment.
5.
Allow 20 minutes for the equipment to
warm up.
6.
Photocopy the test record and use it to
record the test results. (See page 48.)
39
P6021A Current Probe Instruction Manual