Adjustments
Midband Accuracy
This procedure describes the midband accuracy adjustments for the probe. You must complete the
Midband Accuracy Test
in the
Performance Veri
fi
cation
procedure before making this adjustment. Keep the test setup intact to perform the adjustment.
1.
Set the calibrator output to 2 V
RMS
@ 50 kHz.
2.
Set the digital multimeter (DMM) to measure AC volts,
≥
2.0 V range.
3.
Set the probe to the 2 mA/mV range.
4.
Measure the output amplitude on the DMM.
5.
Adjust the 2 mA/mV control to display 20.0 mV,
±0.6 mV on the DMM.
6.
Set the probe to the 10 mA/mV range.
7.
Adjust the 10 mA/mV control to display 4.0 mV,
±0.12 mV on the DMM.
8.
Disconnect the probe from the calibrator and disable
the output.
53
P6021A Current Probe Instruction Manual
Содержание P6021A
Страница 1: ...x P6021A 60 MHz 15 Amp AC Current Probe ZZZ Instruction Manual P071300400 071 3004 00...
Страница 2: ......
Страница 3: ...P6021A 60 MHz 15 Amp AC Current Probe ZZZ Instruction Manual xx www tektronix com 071 3004 00...
Страница 6: ......
Страница 10: ...Table of Contents iv P6021A Current Probe Instruction Manual...
Страница 52: ...Specifications Figure 2 Typical insertion impedance versus frequency 31 P6021A Current Probe Instruction Manual...
Страница 53: ...Specifications Figure 3 Maximum peak pulse versus pulse width P6021A Current Probe Instruction Manual 32...
Страница 56: ...Specifications Figure 4 Mechanical dimensions 35 P6021A Current Probe Instruction Manual...
Страница 73: ...Adjustments 3 Remove the two halves and then remove the metal shield P6021A Current Probe Instruction Manual 52...
Страница 78: ...Maintenance 57 P6021A Current Probe Instruction Manual...