![Tektronix P6021A Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/tektronix/p6021a/p6021a_instruction-manual_1078081059.webp)
Performance Veri
fi
cation
Equipment Required
Table 6 lists the equipment required for the performance veri
fi
cation procedure.
Table 6: Test equipment
Description and quantity
Performance requirement
Recommended example
1
Oscilloscope
TekVPI interface, 500 MHz or greater bandwidth
Tektronix DPO4000
Calibration generator
ACA: 0 to ±1 A, 50 Hz, square wave output
ACV 0 to 2 V
RMS
, 50 kHz, sine wave output
Rise time
≤
1 ns
Fluke 9100
Digital multimeter (DMM)
≥
2 volt range,
≥
1% accuracy
Keithley 2700
HF current loop
50
Ω
±0.5%, BNC male connector
067-2396-xx
BNC cable (2)
50
Ω
, 0.76 m (30 in) length
012-0117-xx
Test lead
Banana plugs at both ends, lead diameter not to
exceed 5 mm (0.2 in)
012-0039-00
Adapter (2)
BNC-to-dual banana
103-0090-xx
Adapter
BNC-T
103-0030-xx
Adapter
SMA male-to-BNC female
015-1018-xx
1
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers
P6021A Current Probe Instruction Manual
38
Содержание P6021A
Страница 1: ...x P6021A 60 MHz 15 Amp AC Current Probe ZZZ Instruction Manual P071300400 071 3004 00...
Страница 2: ......
Страница 3: ...P6021A 60 MHz 15 Amp AC Current Probe ZZZ Instruction Manual xx www tektronix com 071 3004 00...
Страница 6: ......
Страница 10: ...Table of Contents iv P6021A Current Probe Instruction Manual...
Страница 52: ...Specifications Figure 2 Typical insertion impedance versus frequency 31 P6021A Current Probe Instruction Manual...
Страница 53: ...Specifications Figure 3 Maximum peak pulse versus pulse width P6021A Current Probe Instruction Manual 32...
Страница 56: ...Specifications Figure 4 Mechanical dimensions 35 P6021A Current Probe Instruction Manual...
Страница 73: ...Adjustments 3 Remove the two halves and then remove the metal shield P6021A Current Probe Instruction Manual 52...
Страница 78: ...Maintenance 57 P6021A Current Probe Instruction Manual...