Con
fi
guring the OM4000 user interface (OUI)
Non-VISA oscilloscope connections (Scope Service Utility)
As mentioned above, the other choice for connecting to the oscilloscope and
collecting data is the Scope Service Utility (SSU). The SSU is a program that runs
on each oscilloscope to be connected to the OUI.
NOTE.
The Scope Service Utility runs on the target oscilloscope. Be sure to
install the proper version of SSU for either real-time or equivalent-time (ET)
oscilloscopes. See installation guide.
Once the SSU is installed on the oscilloscope, start the “Socket Server” and the
TekScope oscilloscope application before starting the SSU using the desktop icon.
The Scope Service has a small User Interface shown below.
NOTE.
It is best to have the oscilloscope in single-acquisition mode (not Run
mode). The Scope Service Utility takes data directly from the oscilloscope memory
and serves it up over a WCF interface to the OUI.
OM4000D Series Coherent Lightwave Signal Analyzer
59
Содержание OM4006D
Страница 1: ...xx OM4106D and OM4006D Coherent Lightwave Signal Analyzer ZZZ User Manual P071316002 071 3160 02...
Страница 2: ......
Страница 6: ......
Страница 22: ...Compliance information xvi OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 24: ...Preface xviii OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 100: ...Taking measurements 76 OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 132: ...Appendix A MATLAB variables used by core processing 108 OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 146: ...Appendix D Automatic receiver deskew 122 OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 176: ...Appendix F Configuring two Tektronix 70000 series oscilloscopes 152 OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 202: ...Appendix H Cleaning and maintenance 178 OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 205: ...Index W Waveform averaging 48 OM4000D Series Coherent Lightwave Signal Analyzer 181...