Getting started
12.
On the PC, start the
LRCP
program. (See page 19,
Control Panel (LRCP) user interface
13.
Enter password
1234
when requested.
14.
Select
Con
fi
guration > Device Setup
from the menu to open the Device
Setup window.
15.
Click the
Auto Con
fi
gure
button. LRCP lists the OM-series instrument
connected to the PC. If LRCP does not list the connected instrument, verify
that you entered a correct IP address into the PC and your Ethernet cable is
good. If the IP address was entered correctly, you may need to connect the
OM instrument to a DHCP network to determine if the IP address you used to
set the computer was correct.
16.
(optional) Use the
Friendly Name
fi
eld to create a custom label for each
instrument. There is no limit to the size of the name you enter. Friendly Names
are retained and are associated with the MAC address of each instrument.
17.
Enter the new IP address for the OM instrument in the AutoCon
fi
g screen that
is compatible with your network. For example,
172.17.200.040
.
18.
Edit the Gateway and Net Mask
fi
elds only if necessary (obtain this
information from your network support).
19.
Click
Set IP
.
NOTE.
If you change the instrument to an IP address that is different than
the Subnet of the PC, and click Set IP, the OM series instrument is no longer
detectable or viewable to that PC and LRCP.
20.
Click
OK
to exit the screen and return to the LRCP window.
21.
Exit the LRCP program.
22.
Unplug the network cable from between the PC and the OM instrument.
23.
Connect the OM instrument to the target network switch/router.
24.
Run the LRCP software on the PC connected to the same network as the
OM instrument.
25.
Click
Device Setup
. Click
Auto Con
fi
g
and verify that the instrument is
detected and listed on the display.
12
OM4000D Series Coherent Lightwave Signal Analyzer
Содержание OM4006D
Страница 1: ...xx OM4106D and OM4006D Coherent Lightwave Signal Analyzer ZZZ User Manual P071316002 071 3160 02...
Страница 2: ......
Страница 6: ......
Страница 22: ...Compliance information xvi OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 24: ...Preface xviii OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 100: ...Taking measurements 76 OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 132: ...Appendix A MATLAB variables used by core processing 108 OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 146: ...Appendix D Automatic receiver deskew 122 OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 176: ...Appendix F Configuring two Tektronix 70000 series oscilloscopes 152 OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 202: ...Appendix H Cleaning and maintenance 178 OM4000D Series Coherent Lightwave Signal Analyzer...
Страница 205: ...Index W Waveform averaging 48 OM4000D Series Coherent Lightwave Signal Analyzer 181...