Model 2450 Interactive SourceMeter® Instrument User's Manual
Section 6: Leakage current and insulation resistance
2450-900-01 Rev. A / June 2013
6-9
Setting up the insulation resistance example using the front panel
To set up the application from the front panel:
1. Connect the device under test (DUT) to the rear panel of the Model 2450, as described in
(on page 6-2).
2. Reset the Model 2450.
a. Press the
MENU
key.
b. Under System, select
Manage
.
c. Select
System Reset
, and then select
OK
.
3. Press the TERMINAL FRONT/REAR switch to set the instrument to use the rear-panel terminals
(R should be displayed to the left of the switch).
4. Press the
HOME
key.
5. Press the
FUNCTION
key.
6. Under Source Voltage and Measure, select
Resistance
.
7. Select the button next to Source (at the bottom of the screen).
8. Enter 20 V (or other applicable test voltage) and select
OK
.
9. Press the
OUTPUT ON/OFF
switch to enable the output and start making measurements. The
resistance measurements appear in the measure area (top half) of the home screen.
10. Press the
OUTPUT ON/OFF
switch to turn off the output when you are done taking
measurements.
Viewing the measurements on the front-panel graph
To view the insulation resistance measurements on the front-panel graph while the output is on:
1. Press the
MENU
key.
2. Under Views, select
Graph
.