Section 2: Using the front-panel interface
Model 2450 Interactive SourceMeter® Instrument User's Manual
2-4
2450-900-01 Rev. A / June 2013
To turn a Model 2450 on and off:
1. Disconnect any devices under test (DUTs) from the Model 2450 before turning the instrument on.
2. To turn your instrument on, press the front-panel
POWER
switch to place it in the on (I) position.
A status bar is displayed as the instrument powers up. The Home screen is displayed when
power up is complete.
3. To turn your instrument off, press the front-panel
POWER
switch to place it in the off (O) position.
Turning the Model 2450 output on
You can turn the Model 2450 output on from the front panel and by sending remote commands.
To turn the output on using the front panel:
Press the
OUTPUT ON/OFF
switch. The instrument is in the output-on state when the switch is
illuminated. The instrument is in the output-off state when the switch is not illuminated.
To turn the output on using SCPI commands:
:OUTPut[:STATe] ON
To turn the output on using TSP commands:
smu.source.output = smu.ON
Turning the Model 2450 output off
Turning the Model 2450 output off does not place the instrument in a safe state (an interlock
is provided for this function).
Hazardous voltages may be present on all output and guard terminals. To prevent electrical
shock that could cause injury or death, never make or break connections to the Model 2450
while the instrument is powered on. Turn off the equipment from the front panel or
disconnect the main power cord from the rear of the Model 2450 before handling cables.
Putting the equipment into an output-off state does not guarantee that the outputs are
powered off if a hardware or software fault occurs.
When the source of the instrument is turned off, it may not completely isolate the instrument from the
external circuit. You can use the output-off setting to place the Model 2450 in a known, noninteractive
state during idle periods, such as when changing the device under test. The output-off states that can
be selected for a Model 2450 are normal, high-impedance, zero, or guard.
See "Output-off state" in the Reference Manual for additional details.
Using the front panel:
Press the
OUTPUT ON/OFF
switch. The instrument is in the output-on state when the switch is
illuminated. The instrument is in the output-off state when the switch is not illuminated.
Using SCPI commands:
To turn the output on, send the command:
:OUTPut[:STATe] ON
To turn the output off, send the command:
:OUTPut:STATe OFF