2260B Series User Manual
88
Note
Ensure load cables have sufficient
current capacity.
Page 44
Re-attach the protection dummy plug when not in
use.
Master-Slave Series Operation
Master-Slave
Configuration
Before using the power supplies in series, the
master and slave units need to be configured.
1.
Configure the OVP and OCP
settings for the master unit.
Page 60
2.
For each unit, hold the Function
key while turning the power on to
enter the power on configuration
settings.
3.
Configure F-93 (Master/Slave)
setting for each master/slave unit.
Page 120
Unit
F-93
Master (local or series operation)
0
Slave unit (series)
4
4.
Cycle the power on the units (reset the power).
Note
Configuration settings can be checked for both the
master and slave units by pressing the Function
key.