GETTING STARTED
37
Grounded output
terminal
If the positive or negative terminal is connected
to the protective ground terminal, the
insulation capacity needed for the load and
load cables is greatly reduced. The insulation
capacity only needs to be greater than the
maximum output voltage of the power supply
with respect to ground.
2260B
Load
Ext-V
Ext-R
Analog
connector
( ) Insulation capacity
≥
voltage of power
supply with respect to ground
CAUTION
If using external voltage control, do not ground
the external voltage terminal as this will create a
short circuit.
Содержание Keithley 2260B Series
Страница 5: ...Series 2260B Multi Range Programmable DC Power Supplies USER MANUAL ISO 9001 CERTIFIED MANUFACTURER ...
Страница 14: ...2260B Series User Manual 10 Considerations 33 Grounding 36 ...
Страница 43: ...OPERATION 39 Remove Test Sequence 99 Test Sequence Error Messages 100 ...
Страница 52: ...2260B Series User Manual 48 Removal Reverse the procedure to remove the terminal covers ...
Страница 111: ...CONFIGURATION 107 Note Power On and Calibration settings can only be set during power up ...