background image

For safety, instruments and accessories must be used in accordance with the operating instructions. If the instruments or 

accessories are used in a manner not specified in the operating instructions, the protection provided by the equipment may be 

impaired. 
Do not exceed the maximum signal levels of the instruments and accessories. Maximum signal levels are defined in the 

specifications and operating information and shown on the instrument panels, test fixture panels, and switching cards. 
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard. 
Chassis connections must only be used as shield connections for measuring circuits, NOT as protective earth (safety ground) 

connections. 
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use 

of a lid interlock. 

If a 

  screw is present, connect it to protective earth (safety ground) using the wire recommended in the user documentation. 

The 

  symbol on an instrument means caution, risk of hazard. The user must refer to the operating instructions located in the 

user documentation in all cases where the symbol is marked on the instrument. 

The 

  symbol on an instrument means warning, risk of electric shock. Use standard safety precautions to avoid personal 

contact with these voltages. 

The 

  symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns. 

The 

  symbol indicates a connection terminal to the equipment frame. 

If this 

  symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be 

properly disposed of according to federal, state, and local laws. 
The 

WARNING

 heading in the user documentation explains hazards that might result in personal injury or death. Always read 

the associated information very carefully before performing the indicated procedure. 
The 

CAUTION

 heading in the user documentation explains hazards that could damage the instrument. Such damage may 

invalidate the warranty. 

The 

CAUTION

 heading with the 

  symbol in the user documentation explains hazards that could result in moderate or minor 

injury or damage the instrument. Always read the associated information very carefully before performing the indicated 

procedure. Damage to the instrument may invalidate the warranty.   
Instrumentation and accessories shall not be connected to humans. 
Before performing any maintenance, disconnect the line cord and all test cables. 
To maintain protection from electric shock and fire, replacement components in mains circuits — including the power 

transformer, test leads, and input jacks — must be purchased from Keithley. Standard fuses with applicable national safety 

approvals may be used if the rating and type are the same. The detachable mains power cord provided with the instrument may 

only be replaced with a similarly rated power cord. Other components that are not safety-related may be purchased from other 

suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through 

Keithley to maintain accuracy and functionality of the product). If you are unsure about the applicability of a replacement 

component, call a Keithley office for information. 
Unless otherwise noted in product-specific literature, Keithley instruments are designed to operate indoors only, in the following 

environment: Altitude at or below 2,000 m (6,562 ft); temperature 0 °C to 50 °C (32 °F to 122 °F); and pollution degree 1 or 2. 
To clean an instrument, use a cloth dampened with deionized water or mild, water-based cleaner. Clean the exterior of the 

instrument only. Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that 

consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer) should never 

require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board 

should be returned to the factory for proper cleaning/servicing. 
Safety precaution revision as of June 2017. 

Содержание KEITHLEY 2100

Страница 1: ...Calibration Manual 2100 905 01 Rev B April 2020 P2100 905 01B 2100 905 01B www tek com keithley Model 2100 6 1 2 Digit Resolution Digital Multimeter...

Страница 2: ...ct names are trademarks or registered trademarks of Keithley Instruments LLC Other brand names are trademarks or registered trademarks of their respective holders The Lua 5 0 software and associated d...

Страница 3: ...d to mains These instruments will be marked as category II or higher Unless explicitly allowed in the specifications operating manual and instrument labels do not connect any instrument to mains Exerc...

Страница 4: ...lains hazards that could damage the instrument Such damage may invalidate the warranty The CAUTION heading with the symbol in the user documentation explains hazards that could result in moderate or m...

Страница 5: ...nction adjustment DC and AC 4 1 Voltage function adjustment 4 1 DC voltage range adjustment 4 2 AC voltage range adjustment 4 4 Ohms function adjustment 5 1 Ohms function adjustment 5 1 Resistance low...

Страница 6: ...n also call the Tektronix corporate headquarters toll free inside the U S and Canada only at 1 800 833 9200 For worldwide contact numbers visit tek com contact us Introduction This manual introduces t...

Страница 7: ...the Model 2100 To install the USB device drive 1 Install the USB device driver and KI Tool Software from the Keithley Instruments Model 2100 Product Information CD 2 Connect the 2100 to the USB port...

Страница 8: ...onsiderations When performing the calibration procedure Make sure the equipment is properly warmed up and connected to the appropriate input jacks Also make sure the correct input jacks are selected w...

Страница 9: ...f your 2100 has firmware versions 1 03 or 1 04 the firmware must be updated to the latest version before this procedure can be performed To identify the firmware version press the QUERY button and ent...

Страница 10: ...il 2020 1 5 Calibration code To unlock the calibration function 1 Unlock the calibration function by sending the SCPI command CAL PROT CODE up to 8 character string The default code command is CAL PRO...

Страница 11: ...ze for a minimum of five minutes before proceeding Make sure to connect the low thermal short properly to the HI LO and SENSE terminals Keep drafts away from low thermal connections to avoid thermal d...

Страница 12: ...es before proceeding Make sure to connect the low thermal short properly to the HI LO and SENSE terminals Keep drafts away from low thermal connections to avoid thermal drift which could affect calibr...

Страница 13: ...s are 100 mV 1 V 10 V 100 V and 750 V RMS at 1 kHz Do not apply more than 1000 V peak to the 2100 multimeter Applying excess voltage may damage the meter or cause electric shock resulting in personal...

Страница 14: ...8 Adjust the full scale measurement by sending the following SCPI command CAL PROT DC STEP 2 10 9 Wait until the following message is displayed on the 2100 Cali OK 10 Set the 5700A output mode to STAN...

Страница 15: ...1 100 30 Wait until the following message is displayed on the 2100 Cali OK 31 Set the 5700A output to 100 V dc 32 Adjust the full scale measurement by sending the following SCPI command 33 heaWait unt...

Страница 16: ...1 V 8 Adjust the full scale measurement by sending the following SCPI command CAL PROT AC STEP 2 0 1 9 Wait until the following message is displayed on the 2100 Cali OK 10 Set the 5700A output mode to...

Страница 17: ...displayed on the 2100 Cali OK 31 Set the 5700A output to 100 V 32 Adjust the full scale measurement by sending the following SCPI command CAL PROT AC STEP 2 100 33 Wait until the following message is...

Страница 18: ...ize for a minimum of five minutes before proceeding 3 Set the 2100 to the two wire 1 M resistance function by sending the following SCPI command CONF RES 1E6 4 Set the measurement rate to 10 PLC by se...

Страница 19: ...ate to 10 PLC by sending the following SCPI command FRES NPLC 10 17 Set the average function to 100 counts by sending the following SCPI command AVERAGE COUNT 100 18 Set the 5700A output mode to OPERA...

Страница 20: ...100 Cali OK 40 Set the 5700A output to 10 k 41 Adjust the full scale measurement by sending the following SCPI command CAL PROT DC STEP 2 ACTUAL CALIBRATOR OUTPUT 42 Wait until the following message i...

Страница 21: ...and output to STANDBY mode 2 Using shielded low thermal EMF cables connect the Fluke 5700A output HI output LO sense HI and sense LO to the 2100 input HI input LO sense HI and sense LO terminals Allow...

Страница 22: ...E6 20 Wait until the following message is displayed on the 2100 Cali OK 21 Set the 5700A output to 10 M 22 Adjust the full scale measurement by sending the following SCPI command CAL PROT DC STEP 2 AC...

Страница 23: ...of five minutes before proceeding 3 Configure 2100 to 0 01 A dc current range by sending the following SCPI command CONF CURR DC 0 01 4 Set the measurement rate to 10 PLC by sending the following SCP...

Страница 24: ...output to 0 12 A dc 21 Adjust the full scale measurement by sending the following SCPI command CAL PROT DC STEP 2 0 12 22 Wait until the following message is displayed on the 2100 Cali OK 23 Set the 5...

Страница 25: ...ge by sending the following SCPI command CAL PROT AC STEP 2 1 9 Wait until the following message is displayed on the 2100 Cali OK 10 Set the 5700A output to 0 01 A at 1 kHZ 11 Adjust 1 of the range by...

Страница 26: ...example if the calibration date is 5 Jan 2020 send the following SCPI command CAL PROT DATE 05 01 2020 2 Save the calibration constants in EEPROM by sending the following SCPI command CAL PROT SAVE 3...

Страница 27: ...DC voltage accuracy verification Range Test point Limits of accuracy Lower limit Upper limit 100 mV 100 0000 mV 99 9905 mV 100 0095 mV 100 0000 mV 100 0095 mV 99 9905 mV 1 V 1 000000 V 0 999947 V 1 0...

Страница 28: ...V 35 kHz 219 24 V 220 76 V NOTE Verify this point instead of 700 V if a Fluke 5725A is not available Specifications valid after 2 hour warm up a Slow ac filter 3 Hz bandwidth b Pure sine wave input gr...

Страница 29: ...c set for continuous trigger operation b Input bias current 30 pA at 25 C c Input protection 1000 V all ranges two wire input d Measurement rate set to 1 PLC AC current accuracy verification Range Tes...

Страница 30: ...names are the property of Keithley Instruments All other trademarks and trade names are the property of their respective companies Keithley Instruments Corporate Headquarters 28775 Aurora Road Clevel...

Отзывы: