In this section:
Performance verification .......................................................... 8-1
Performance verification
After the adjustment is completed, the 2100 must be verified to make sure that its performance is
within manufacturer’s specifications. Also, specifications are valid after a 2-hour warm up.
Perform accuracy verifications using Fluke 5700A and 5725A standards and a function generator
capable of sourcing frequency with an accuracy of 25 ppm or better. Refer to the following tables for
limits of accuracy on specified ranges.
DC voltage accuracy verification
Range
Test point
Limits of accuracy
Lower limit
Upper limit
100 mV
100.0000 mV
99.9905 mV
100.0095 mV
-100.0000 mV
-100.0095 mV
-99.9905 mV
1 V
1.000000 V
0.999947 V
1.000053 V
-1.000000 V
-1.000053 V
-0.999947 V
10 V
10.00000 V
9.99956 V
10.00044 V
-10.00000 V
-10.00044 V
-9.99956 V
100 V
100.0000 V
99.9943 V
100.0057 V
-100.0000 V
-100.0057 V
-99.9943 V
1000 V
1000.000 V
999.935 V
1000.065 V
-1000.000 V
-1000.065 V
-999.935 V
Specifications valid after 2-hour warm up:
a. A dc set for continuous trigger operation.
b. Input bias current: <30 pA at 25 °C.
c. Input protection: 1000 V all ranges (two-wire input).
d. Measurement rate set to 1 PLC.
Section 8
Performance verification