Model 2100 6½-Digit Resolution Digital Multimeter
Section 5: Ohms function adjustment
2100-905-01 Rev. B April 2020
5-3
30. Wait until the following message is displayed on the 2100:
Cali OK
31.
Set the 5700A output to 100 kΩ.
32. Adjust the full-scale measurement by sending the following SCPI command:
CAL:PROT:DC:STEP 2,<ACTUAL CALIBRATOR OUTPUT>
33. Wait until the following message is displayed on the 2100:
Cali OK
34. Set the 5700A output mode to
STANDBY
.
35. Set the 2100 to the four-
wire, 10 kΩ resistance function by sending the following SCPI command:
CONF:FRES 1E4
36. Set the measurement rate to 10 PLC by sending the following SCPI command:
FRES:NPLC 10
37.
Set the 5700A to 0Ω a
nd output mode to
OPERATE
.
38. Adjust the input offset by sending the following SCPI command:
CAL:PROT:DC:STEP 1,1E4
39. Wait until the following message is displayed on the 2100:
Cali OK
40.
Set the 5700A output to 10 kΩ.
41. Adjust the full-scale measurement by sending the following SCPI command:
CAL:PROT:DC:STEP 2,<ACTUAL CALIBRATOR OUTPUT>
42. Wait until the following message is displayed on the 2100:
Cali OK
43. Set the 5700A output mode to
STANDBY
.
44. Set the 2100 to the four-
wire, 1 kΩ resistance function by sending the follow
ing SCPI command:
CONF:FRES 1E3
45. Set measurement rate to 10 PLC by sending the following SCPI command:
FRES:NPLC 10
46.
Set the 5700A to 0Ω and output
mode to
OPERATE
.
47. Adjust the input offset by sending the following SCPI command:
CAL:PROT:DC:STEP 1,1E3
48. Wait until the following message is displayed on the 2100:
Cali OK
49.
Set the 5700A output to 1 kΩ.
50. Adjust the full-scale measurement by sending the following SCPI command:
CAL:PROT:DC:STEP 2,<ACTUAL CALIBRATOR OUTPUT>
51. Wait until the following message is displayed on the 2100:
Cali OK
52. Set the 5700A output mode to
STANDBY
.
53. Set the 2100 to the four-
wire, 100 Ω resistance function by sending the following SCPI
command:
CONF:FRES 1E2