CH<x>:PROBECOntrol
This command sets or queries multirange probe range-control policy preference of the probe that is attached to CH<x>. The
channel number is specified by x.
Syntax
CH<x>:PROBECOntrol {AUTO|MANual}
CH<x>:PROBECOntrol?
Arguments
AUTO
sets the values. The probe range is automatically calculated.
MANual
allows you to select various valid values for the probe connected to a
particular channel.
Examples
CH2:PROBECONTROL AUTO
sets the values and the probe range is
automatically calculated.
CH2:PROBECONTROL?
might return
MANUAL
indicating that you can select
various valid values for the probe connected to channel 2.
CH<x>:PROBEFunc:EXTAtten
This command is used to specify the attenuation value as a multiplier to the given scale factor on the specified channel. The
channel is specified by x.
The query form of this command returns the user-specified attenuation.
Syntax
CH<x>:PROBEFunc:EXTAtten <NR3>
CH<x>:PROBEFunc:EXTAtten?
Arguments
<NR3>
is the attenuation value, which is specified as a multiplier in the range from
1.00E-10 to 1.00E+10.
Examples
CH1:PROBEFUNC:EXTATTEN 167.00E-3
specifies an external
attenuation, which is connected between the your input signal and the input of the
probe attached to Channel 1.
CH2:PROBEFUNC:EXTATTEN?
might return
1.0000E+00
, indicating that
the probe attached to Channel 2 is connected directly to the user's signal.
CH<x>:PROBEFunc:EXTDBatten
This command sets or queries the input-output ratio (expressed in decibel units) of external attenuation or gain between the
signal and the instrument input channels. The channel is specified by x.
The query form of this command returns the user-specified attenuation in decibels.
Syntax
CH<x>:PROBEFunc:EXTDBatten <NR3>
CH<x>:PROBEFunc:EXTDBatten?
Arguments
<NR3>
is the attenuation value, which is specified in the range from -200.00 dB to
200.00 dB.
Examples
CH3:PROBEFUNC:EXTDBATTEN 2.5
specifies an external 2.5 dB
attenuator on Channel 3.
CH1:PROBEFUNC:EXTDBATTEN?
might return
2.5000E+00
, indicating
that the attenuation for Channel 1 is 2.5 dB.
CH<x>:PROBEFunc:EXTUnits
This command sets the unit of measurement for the external attenuator of the specified channel. The channel is specified by
x. The alternate units are used if they are enabled. Use the
CH<x>:PROBEFunc:EXTUnits:STATE
command to
enable or disable the alternate units.
Remote programming
TIVP Series IsoVuTM Measurement System User Manual
54