CH<x>:PRObe:FORCEDRange

The command selects the dynamic range of probe (1 of 9) in +/-V. It is dependent on the attached probe tip. The channel is
specified by x. The command should only be used when 

CH<x>:PROBECONTROL

 is set to 

MANUAL

.

Table 12: Probe tip cables and dynamic ranges

Probe tip cable

Dynamic Range +/-V

No tip or 1X tip

0.02 | 0.04 | 0.08 | 0.16 | 0.32 | 0.64 | 1.28 | 2.56 | 5.0

10X

0.2 | 0.4 | 0.8 | 1.6 | 3.2 | 6.4 | 12.8 | 25.6 | 50.0

50X

1.0 | 2.0 | 4.0 | 8.0 | 16.0 | 32.0 | 64.0 | 128.0 | 250.0

100X

2.0 | 4.0 | 8.0 | 16.0 | 32.0 | 64.0 | 128.0 | 256.0 | 500.0

500X

10.0 | 20.0 | 40.0 | 80.0 | 160.0 | 320.0 | 640.0 | 1280.0 |
2500.0

The query returns the dynamic range of the probe tip in +/-V.

Syntax

CH2:PRObe:FORCEDRange <NR3>
CH2:PRObe:FORCEDRange?

Arguments

<NR3>

 specifies the probe dynamic range

Examples

If a current probe is attached to the Channel 1 input,

CH1:PROBE:FORCEDRANGE 5.0

 sets the attached probe to its 5 V range.

CH3:PROBE:FORCEDRANGE?

 might return 

5.0000

, indicating that the range

of the probe attached to the Channel 3 is set to 5 V.

CH<x>:PRObe:GAIN? (Query Only)

The command returns the gain factor of the currently selected range (inverse of attenuation). The channel is specified by x.

Syntax

CH<x>:PRObe:GAIN?

Examples

CH2:PROBE:GAIN?

 might return 

100.0000E-3

, indicating that the attached

10X probe delivers 0.1 V to the Channel 2 BNC for every 1.0 V applied to the probe
input.

CH<x>:PRObe:ID? (Query Only)

This query-only command returns the type and serial number of the probe that is attached to the specified channel. The
channel is specified by x.

Syntax

CH<x>:PRObe:ID?

Examples

CH2:PROBE:ID?

 might return 

"B010289";"TIVP1"

, indicating that a

TIVP1 probe with serial number B010289 is attached to Channel 2.

Remote programming

TIVP Series  IsoVuTM Measurement System User Manual 

51

Содержание IsoVu TIVP Series

Страница 1: ...TIVP Series IsoVuTM Measurement System User Manual P 071369200 071 3692 00...

Страница 2: ......

Страница 3: ...VP Series IsoVuTM Measurement System User Manual Supports MSO 4 5 6 Series Firmware version 1 28 and above Register now Click the following link to protect your product www tek com register 071 3692 0...

Страница 4: ...ued and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of...

Страница 5: ...cation 20 Product description 21 Compensation box 21 Sensor head 21 Sensor tip cables 21 Models 22 Supported oscilloscopes 22 Accessories 22 Standard accessories 22 Optional accessories 23 Getting sta...

Страница 6: ...e sensor tip cables 41 Probe tip tripod 41 Installing the probe tip adapters 41 Installing the square pins on the circuit board 43 User service 45 Service offerings 45 Warranty repair service 45 Calib...

Страница 7: ...sor head 39 Figure 11 Sensor tip cable 39 Figure 12 Compensation box 39 Figure 13 Block diagram 40 Figure 14 Square pin series sensor tip cables 41 Figure 15 Connecting to an adapter on the circuit bo...

Страница 8: ...quired equipment for performance verification 45 Table 6 Test record table 47 Table 7 STATUS LED descriptions 48 Table 8 SELF CAL LED descriptions 48 Table 9 INPUT LED descriptions 48 Table 10 ALT Mod...

Страница 9: ...r products returned to any other locations This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and care Tektronix shall not be ob...

Страница 10: ...IRE LAWFULLY OR THAT IS NOT A LEGITIMATE AUTHORIZED COPY OF TEKTRONIX S SOFTWARE THIS AGREEMENT EXPRESSLY EXCLUDES ANY RIGHTS CONCERNING SUCH ILLEGITIMATE COPIES IF THESE TERMS ARE NOT ACCEPTABLE THE...

Страница 11: ...e Software or Equipment designed to control the manner in which they are used or harvest or mine Tektronix s proprietary content or information from the Software or Equipment THE SOFTWARE MAY NOT BE U...

Страница 12: ...E USE OF THE SOFTWARE IN CONJUNCTION WITH OTHER PRODUCTS OR SERVICES INCLUDING THIRD PARTY PRODUCTS OR SERVICES OR 5 INDEMNIFICATION OF YOU OR OTHERS FOR COSTS DAMAGES OR EXPENSES RELATED TO THE GOODS...

Страница 13: ...ith respect to its use and do not undertake to furnish any support or information relating thereto GENERAL Unless the Customer is the United States Government this Agreement contains the entire agreem...

Страница 14: ...IN CONTRACT STRICT LIABILITY OR TORT INCLUDING NEGLIGENCE OR OTHERWISE ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE http www freescale com mail...

Страница 15: ...Open License WOL Permission to use copy modify distribute and sell this software and its documentation for any purpose is hereby granted without fee provided that the above copyright notice and this l...

Страница 16: ...ers supplied with the product or indicated by Tektronix to be suitable for the product Observe all terminal ratings To avoid fire or shock hazard observe all rating and markings on the product Consult...

Страница 17: ...nergize the test circuit before connecting or disconnecting the sensor tip cable Terms in this manual These terms may appear in this manual Warning Warning statements identify conditions or practices...

Страница 18: ...ation and Requirements Edition 3 2014 US 21CFR PT1010 Performance Standard for Electronic Parts 2015 US 21CFR PT1040 Performance Standards for Light Emitting Products 2015 Equipment type Test and meas...

Страница 19: ...on provides information about the environmental impact of the product Product end of life handling Observe the following guidelines when recycling an instrument or component Equipment recycling Produc...

Страница 20: ...ejection ratio at DC 140 dB 10 million to 1 at 1 MHz 120 dB 1 million to 1 at 100 MHz and 80 dB 10 000 to 1 at 1 GHz Differential voltages up to 2500 Vpk sensor tip cable dependent Large common mode v...

Страница 21: ...The buttons and indicators on the comp box provide a means for controlling the probe and indicating the overall status Sensor head The sensor head provides an interface between the device under test D...

Страница 22: ...following table list the accessories that are shipped with the probe Accessory Description Part number The 10X MMCX Tip is included in every TIVP The MMCX tip is recommended for the best bandwidth and...

Страница 23: ...nnector TIVPWS500X 1X MMCX sensor tip cable TIVPMX1X Square Pin to MMCX Adapter 0 062 Spacing 131 9677 xx Probe Tip Tripod Support 352 1170 xx Lead MMCX to IC Grabber 196 3546 xx Lead Square Pin to IC...

Страница 24: ...optic cable smaller than 5 inches 12 7 cm Do not twist the fiber optic cable twisting the cable will stress the optical fibers Do not allow kinks or knots to develop in the fiber optic cable Avoid put...

Страница 25: ...midity Operating Comp box 5 to 85 RH Relative Humidity at up to 40 C non condensing Probe head 5 to 85 RH Relative Humidity at up to 40 C 5 to 45 RH above 40 C up to 50 C non condensing Tip Cable Adap...

Страница 26: ...as recommended in this document Do not access the RF Burn Hazard Zone while taking measurements on a live circuit While measuring high frequency common mode signals there is a risk of RF burns Refer t...

Страница 27: ...termination For more information on the state of the LED see Table 9 INPUT LED descriptions on page 48 4 MENU button Press to open the Probe Vertical menu on the oscilloscope 5 ALT mode indicator Dis...

Страница 28: ...void possible damage to the equipment do not connect the coaxial common shield of the sensor tip cable or SMA input to the high impedance portion of a circuit The additional capacitance can cause circ...

Страница 29: ...erform the self calibration before connecting the tip cable to the DUT and taking the measurement 5 Connect the sensor tip cable end to the DUT If you are using MMCX sensor tip cable connect it to an...

Страница 30: ...he self calibration 1 Connect probe to an oscilloscope channel 2 Press the SELF CAL button on the probe compensation box The Self Cal indicator LED will blink while self calibration is in process 3 Se...

Страница 31: ...to start and monitor self calibration and a status query to determine if self calibration is needed AutoZero AutoZero and self calibration work on different parts of the measurement system Self calibr...

Страница 32: ...are selectable when Range Mode is set to Manual The recommended V div settings are displayed in the table below The ranges shown are for the probe SMA input and 1X tip Multiply the range and V div se...

Страница 33: ...tip cable Start at the top of the table and work down Choose the first tip that meets all of your criteria Table 3 Sensor tip cable selection Sensor tip cable Most sensitive V div setting Dynamic ran...

Страница 34: ...l input range is limited and depends on the tip attenuation the gain range selected and the applied offset The input voltage conditions are divided into several different input ranges Common mode volt...

Страница 35: ...ent Performance verification procedures on page 45 Overview Characteristic TIVP1 TIVP05 TIVP02 Bandwidth 1 GHz 500 MHz 200 MHz Rise time 450ps 850ps 2ns Differential Input Voltage Range Offset Range D...

Страница 36: ...ain accuracy 2 System noise rms Sensor tip cable 20 mV range most sensitive 320 mV range 5 V range widest range SMA Input 50 mode 0 43 mV rms 1 46 mV rms 48 mV rms SMA Input 1 M mode 0 43 mV rms 1 46...

Страница 37: ...ure shows the typical CMRR values for the supported sensor tip cables Note The values for the TIVPMX50X TIVPSQ100X and TIVPWS500X sensor tip cables are preliminary Figure 7 Typical CMRR values for the...

Страница 38: ...VP physical specifications Table 4 Physical specifications Characteristic Description Net weight Weight does not include accessories and packaging Tip cable assembly 25g 0 055 lb Probe head 0 363 kg 0...

Страница 39: ...Probe dimensions Figure 10 Sensor head Figure 11 Sensor tip cable Figure 12 Compensation box Reference information TIVP Series IsoVuTM Measurement System User Manual 39...

Страница 40: ...F when the sensor head is placed six 6 inches 15 25 cm above a ground plane To minimize the effects of common mode capacitive loading consider the following items Whenever possible choose a reference...

Страница 41: ...shows an example of connecting to an adapter on the circuit board with the tripod it reduces stresses on the test point Figure 15 Connecting to an adapter on the circuit board with the probe tip trip...

Страница 42: ...h down on the adapter to seat it in place on the circuit board Figure 18 Pushing the MMCX to 0 062 inch 1 57 mm adapter in place Figure 19 Pushing the MMCX to 0 1 inch 2 54 mm adapter in place When th...

Страница 43: ...to remove the spacer as shown in the figure Figure 21 Removing the header from square pins on the circuit board Tektronix provides a set of solder pins 0 018 inch 0 46 mm diameter to install on the c...

Страница 44: ...solder pins into the soldering aide as shown in the following illustration Figure 22 Using the soldering aide to install the square pins on the circuit board 2 Use the soldering aide to hold the squa...

Страница 45: ...ntenance CAUTION To prevent damage to the measurement system do not expose it to sprays liquids or solvents Avoid getting moisture inside the comp box or sensor head when cleaning the exterior Clean t...

Страница 46: ...up action 2 Run Signal Path Compensation if recommended in Utility Calibration 3 Run the self calibration routine Self calibration on page 30 4 Enable the TIVP Channel and use the following Vertical m...

Страница 47: ...n procedures Table 6 Test record table Model number Serial number Procedure performed by Date System RMS noise performance check Probe Range Limit Test result TIVP1 20 mV 1 mV RMS TIVP05 20 mV 0 7 mV...

Страница 48: ...may be required Red Blinking Self Test failure No Power to probe head Try unplugging and plugging back in Service of probe may be required Table 8 SELF CAL LED descriptions LED Status Green Solid Sel...

Страница 49: ...lower noise There is no signal detected the waveform is a flat line Remove the sensor tip cable and check its continuity referencing the input impedance table Try putting a signal directly into the pr...

Страница 50: ...el is specified by x Syntax CH x PRObe Examples CH2 PROBE might return 1 0000E 01 RESISTANCE 1 0000E 07 UNITS V ID TYPE 10X SERNUMBER N A for a 10X probe indicating that among other parameters the att...

Страница 51: ...attached to the Channel 1 input CH1 PROBE FORCEDRANGE 5 0 sets the attached probe to its 5 V range CH3 PROBE FORCEDRANGE might return 5 0000 indicating that the range of the probe attached to the Cha...

Страница 52: ...ce of the probe that is attached to the specified channel The channel is specified by x Syntax CH x PRObe RESistance Examples CH2 PROBE RESISTANCE might return 10 00000E 6 indicating that the input re...

Страница 53: ...us Returns Returns an integer number that represents the sum total of binary error bits B0 B15 The error bits are not displayed they are concatenated into the integer value The following is a list of...

Страница 54: ...fies an external attenuation which is connected between the your input signal and the input of the probe attached to Channel 1 CH2 PROBEFUNC EXTATTEN might return 1 0000E 00 indicating that the probe...

Страница 55: ...ent turns on external units NR1 0 turns off external units any other value turns on external units Examples CH2 PROBEFunc EXTUnits STATE ON turns on external units CH2 PROBEFunc EXTUnits STATE might r...

Страница 56: ...ut offset 33 input specifications 25 installing the probe tip adapters 42 Iowegian ScopeIIR xv IsoVu 20 L laser certification 20 M measurement system setup 46 menu button 32 MMCX connectors 21 models...

Страница 57: ...rs 29 square pin connectors 21 standard accessories 22 supported oscilloscopes 22 T test record 47 TIVP02 22 TIVP05 22 TIVP1 22 troubleshooting 48 U user service 45 V voltage vs frequency derating gra...

Отзывы: