Theory of Operation
2- 8
DSA8200 Digital Serial Analyzer and Modules Service Manual
1
12.5 GHz: The sampler bandwidth and response is set during the calibration
such that the combined system will have an Optical Bandwidth (--6 dB)
>12.5 GHz.
1
20 GHz: The sampler bandwidth and response is set during the calibration
such that the combined system will have an Optical Bandwidth (--6 dB)
>20 GHz.
1
30 GHz: The sampler bandwidth and response is set during the calibration
such that the combined system will have an Optical Bandwidth (--6 dB)
>30 GHz.
The 80C03 module supports conformance testing of both short and long
bandwidth (700-1650 nm) signals at 1.063, 1.250, and 2.488 Mb/s as well as
general purpose testing with >2.3 GHz optical bandwidth. Its amplified optical to
electrical converter design enables the user to examine very low-level optical
signals.
1
OC48: The electrical sampler is adjusted to approximately 12.5 GHz (--3 dB)
bandwidth and signal is not routed through any filter (signal is sent through
the straight-through path). The O/E converter is designed by the vendor to
match the OC48 reference receiver response curve. The span adjustment
provides limited adjustment of the frequency response. It is adjusted at the
factory so that the combined system will have an OC48 (STM-64) Reference
Receiver response. This mode is synonymous with the 2.3 GHz maximum
bandwidth setting.
1
FC1063: The electrical sampler is adjusted and O/E converter span voltage
are the same values as in the OC48 mode. The signal is routed through a
hardware filter designed to result in the combined system having an FC1063
(1.0625 Gb/s Fibre Channel) Reference Receiver response.
1
GBE: The electrical sampler is adjusted and O/E converter span voltage are
the same values as in the OC48 mode. The signal is routed through a
hardware filter designed to result in the combined system having an GBE
(1.25 Gb/s Gigabit Ethernet) Reference Receiver response.
The 80C03 module can be configured with clock recovery that supports Fibre
Channel 1063 (1.063 Gb/s) and OC-48 / STM-16 (2.488 Gb/s) standards.
The 80C04 module is optimized for testing of long wavelength (1100--1650 nm)
signals at either 9.953 Gb/s or 10.664 Gb/s. With its high optical bandwidth
>28 GHz, it is also well suited to general-purpose, high-performance optical
component testing.
1
OC192 or 10.664 Gb/s: The electrical sampler is adjusted to give the desired
response, and signal is not routed through any filter (signal is sent through
the straight-through path). The sampler bandwidth and response is optimized
80C03 Optical Sampling
Module
80C04 Optical Sampling
Module
Содержание DSA8200 Series
Страница 1: ...xx DSA8200 Digital Serial Analyzer and Modules ZZZ Service Manual P071204904 071 2049 04 ...
Страница 2: ......
Страница 6: ......
Страница 18: ...General Safety Summary xii DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Страница 20: ...Service Safety Summary xiv DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Страница 23: ...Operating Information ...
Страница 24: ......
Страница 33: ...Theory of Operation ...
Страница 34: ......
Страница 49: ...Adjustment Procedures ...
Страница 50: ......
Страница 57: ...Maintenance ...
Страница 58: ......
Страница 64: ...Maintenance 4 6 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Страница 138: ...Removal and Installation Procedures 4 80 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Страница 150: ...Repackaging Instructions 4 92 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Страница 151: ...Diagrams ...
Страница 152: ......
Страница 161: ...Replaceable Parts ...
Страница 162: ......
Страница 196: ...Replaceable Parts List 6 34 DSA8200 Digital Serial Analyzer and Modules Service Manual ...