Removal and Installation Procedures
DSA8200 Digital Serial Analyzer and Modules Service Manual
4--63
Thermal cover
Holding clips
T--15 Torxdrive
spacer post
Screw
Figure 4--38: Thermal cover removal
j.
Disconnect J6 ribbon cable connector from the large module circuit
board.
k.
Disconnect J5 coaxial cable connector from the large module circuit
board.
l.
Disconnect the three coaxial cables connected to the front panel;
INTERNAL CLOCK OUTPUT, DC CALIBRATION OUTPUT, AND
EXTERNAL 10 MHz REFERENCE INPUT.
m.
Disconnect the TRIGGER PROBE POWER ribbon connector from J25
on the Electrical front end board.
n.
Slide the Acquisition assembly back toward the rear of the instrument
then lift up and out of the instrument.
4.
Reinstallation:
Do, in reverse order, substeps g through n.
Содержание DSA8200 Series
Страница 1: ...xx DSA8200 Digital Serial Analyzer and Modules ZZZ Service Manual P071204904 071 2049 04 ...
Страница 2: ......
Страница 6: ......
Страница 18: ...General Safety Summary xii DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Страница 20: ...Service Safety Summary xiv DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Страница 23: ...Operating Information ...
Страница 24: ......
Страница 33: ...Theory of Operation ...
Страница 34: ......
Страница 49: ...Adjustment Procedures ...
Страница 50: ......
Страница 57: ...Maintenance ...
Страница 58: ......
Страница 64: ...Maintenance 4 6 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Страница 138: ...Removal and Installation Procedures 4 80 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Страница 150: ...Repackaging Instructions 4 92 DSA8200 Digital Serial Analyzer and Modules Service Manual ...
Страница 151: ...Diagrams ...
Страница 152: ......
Страница 161: ...Replaceable Parts ...
Страница 162: ......
Страница 196: ...Replaceable Parts List 6 34 DSA8200 Digital Serial Analyzer and Modules Service Manual ...