Acquisition
4.
To set channel thresholds used when
converting an analog waveform to digital
format for inclusion in a bus, click the
Digital Properties
button.
5.
To choose whether to share threshold
settings or have independent threshold
settings, click
Channel Settings
and
select
Shared
or
Independent
from the
list.
6.
For each threshold you want to set, click
the channel threshold. Set the threshold
using the keypad or multipurpose knobs.
7.
When the required thresholds have been
set, click the
OK
button.