Acquisition
Starting and Stopping an Acquisition
After the channels that you want to acquire are selected, use the following procedure.
1.
Push the front-panel
RUN/STOP
button
to start the acquisition.
2.
Push the
RUN/STOP
button again to
stop the acquisition.
3.
To take a single acquisition, push the
Single
button.
MSO70000/C, DSA70000B/C, DPO70000B/C, and DPO7000 Series
MSO5000 and DPO5000 Series
Selecting the Horizontal Mode
Your instrument has three horizontal modes. Automatic is the default mode. Select the horizontal mode that works best
for your test setup.
To set the horizontal mode, select
Horiz/Acq
> Horizontal/Acquisition Setup
to display
the horizontal control window. Choose one
of the modes described below.
In Automatic mode you can set the Scale and
Sample Rate. Record length is a dependent
variable. If changing the scale would cause
the record length to exceed the Record
Length Limit, the sample rate is decreased
to the next available setting.
If the sampling mode is real time and
the sample rate is at the real time limit,
attempting to increase the sample rate has
no effect.
40
MSO70000/C, DPO/DSA70000B/C, DPO7000, and MSO/DPO5000 Series User Manual