Performance Veri
fi
cation
c.
Set the
Termination
(input impedance) to
1 M
Ω
.
d.
Repeat steps 9 through 13.
15.
Repeat the procedure for all remaining channels as follows:
a.
Push the front-panel button to deselect the channel that you have already
tested.
b.
Move the DC voltage source connection to the next channel input to be
tested.
c.
Starting from step 6, repeat the procedure for each channel.
MSO4000B and DPO4000B Series Speci
fi
cations and Performance Veri
fi
cation
65