Speci
fi
cations
Table 4: Trigger speci
fi
cations (cont.)
USB
Data Rates Supported:
HS: 480 Mbps, Full: 12 Mbps, Low: 1.5 Mbps
Trigger On:
Sync, Reset, Suspend, Resume, End of Packet, Token (Address) Packet, Data
Packet, Handshake Packet, Special Packet, Error
NOTE.
HIGH SPEED support available only on MSO4104B, DPO4104B, MSO4104B-L,
DPO4104B-L, MSO4102B-L, DPO4102B-L, MSO4102B, and DPO4102B models.
Ethernet
Bit Rate:
10BASE-T, 10 Mbps; 100BASE-TX, 100 Mbps
Trigger On:
Start Frame Delimiter (SFD), MAC Address, MAC Length/Type, IP Header, TCP
Header, TCP/IPv4/MAC Client Data, End of Packet, Idle, FCS (CRC) Error, MAC Q-Tag
control Information.
Digital Acquisition Speci
fi
cations, MSO4000B Series
The following table shows the digital acquisition speci
fi
cations for the MSO4000B
Series oscilloscopes.
Table 5: Digital acquisition speci
fi
cations, MSO4000B Series
Characteristic
Description
Threshold voltage range
-40 V to +40 V
Digital channel timing resolution
2 ns main memory, 60.6 ps for MagniVu memory
Logic threshold accuracy
±(100 mV + 3% of threshold setting after calibration)
Requires valid SPC, as described in step 2 of the
Self Test
. (See page 55,
Self Test
.)
Minimum detectable pulse width,
typical
1 ns
Using MagniVu memory. Requires the use of 342-1140-00 ground clip on each channel.