Performance Veri
fi
cation
Self Test
This procedure uses internal routines to verify that the oscilloscope functions and
passes its internal self tests. No test equipment or hookups are required.
1.
Run the System Diagnostics (may take several minutes):
a.
Disconnect everything from the oscilloscope inputs.
b.
Push the front-panel
Default Setup
button.
c.
Push the
Utility
menu button.
d.
Push the
Utility Page
lower-bezel button.
e.
Turn the
Multipurpose a
knob to select
Self Test
.
f.
Push the
Self Test
lower-bezel button. The Loop X Times side-bezel
menu will be set to
Loop 1 Times
.
g.
Push the
OK Run Self Test
side-bezel button.
h.
Wait. The internal diagnostics perform an exhaustive veri
fi
cation of
proper instrument function. This veri
fi
cation may take several minutes.
i.
Verify that the status of all tests on the readout is
Pass
.
j.
Push the
Menu Off
button twice to clear the dialog box and Self Test
menu.
2.
Run the signal path compensation routine (may take 5 to 15 minutes):
a.
Push the front-panel
Default Setup
button.
b.
Push the
Utility
menu button.
c.
Push the
Utility Page
lower-bezel button.
d.
Turn the
Multipurpose a
knob to select
Calibration
.
e.
Push the
Signal Path
lower-bezel button.
f.
Push the
OK-Compensate Signal Paths
side bezel button.
g.
When the signal path compensation is complete, push the
Menu Off
button twice to clear the dialog box and Self Test menu.
h.
Check the lower-bezel
Signal Path
button to verify that the status is
Pass
.