Setting Jitter/Wander Parameters
3–184
CTS850 SDH/PDH Test Set User Manual
Executing Jitter Calibration
To calibrate the jitter measurement system:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
UTILITY
CALIBRATION
System
Jitter
Measurement
Routine
All
Calibration
Control
Run
To calibrate the jitter generation system:
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
UTILITY
CALIBRATION
System
Jitter Generation
Routine
All
STM–0
STM–1
STM–4
2 Mb/s
34 Mb/s
45 Mb/s
140 Mb/s
Analog output
Calibration
Control
Run
Jitter Generation for all rates takes about 45 minutes to run.
A status message indicates that the calibration is running, has
completed and has passed, or has failed.