Setting Jitter/Wander Parameters
3–174
CTS850 SDH/PDH Test Set User Manual
To set the modulation amplitude (in unit intervals):
Press Menu
Button
Select Menu
Page
Highlight
Parameter
Select Choice
TRANSMIT
JITTER &
WANDER
Jitter Amplitude
300.00 UI (Max)
150.00 UI (Max/2)
0.0 UI
H
Select USER DEFINED to specify a jitter amplitude different
from the predefined choices. Use choices Coarse, Medium or
Fine.
Measure Wander on a 2 MHz clock output from a network element
When generating wander, synchronize the test system to an external
timing source. This can be done by connecting a 2.048 MHz clck to
the rear panel reference timing input and selecting timing source
under TRANSMIT settings. Follow the steps below:
H
Connect 2 MHz reference clock to 2 Mbps/ 2 MHz reference
clock input on Rear Panel of test unit.
H
Connect External 2 MHz clock to be measured to Jitter/ Clock
input on Rear Panel of test unit.
H
Go to TRANSMIT settings. Set the Transmit Clock to 2 MHz
external. The Receive Clock is the Same. No Wander Measure-
ments should be made with Internal Clock.
H
Press Start/Stop twice to Start/Restart the test.
H
For RESULTS, select the Jitter & Wander Screen, and view
Wander Measurements.
Содержание CTS 850
Страница 4: ......
Страница 24: ...Table of Contents xxiv CTS 850 SDH PDH Test Set User Manual...
Страница 33: ...Getting Started...
Страница 34: ......
Страница 49: ...Operating Basics...
Страница 50: ......
Страница 105: ...Reference...
Страница 106: ......
Страница 169: ...Index...
Страница 170: ......
Страница 178: ...Index Index 8 CTS850 SDH PDH Test Set User Manual...