Example 1: Con
fi
gure the Pattern Generator
6.
Touch the CLOCK + icon button and
verify that
Link Clock +/- Signals
is
selected.
This setting links the Clock + and Clock –
outputs together.
7.
Select
Logic Family
in the list and then
select
CML
from the pop-up list.
8.
Repeat the previous two steps to set the
outputs for the Data + output icons.
9.
Touch the
Outputs On/Off
button at the
top of the Generator view to enable the
Clock and Data outputs.
When the Generator outputs are turned
on, the clock and data information
displays under the Clock and Data
buttons.
The Pattern Generator is now transmitting PRBS-7 data at 1 Gb/s on the clock and differential Data outputs set to the
industry standard CML levels.
Tektronix BERTScope Bit Error Rate Analyzers Quick Start User Manual