Adjustment Procedures
AWG7000 Series Service Manual
3-13
b.
Modify the sampling oscilloscope settings as follows:
Vertical External Attenuation: 12 dB
Vertical Scale: 200 mV/div
Horizontal Scale: 50 ps/div
Horizontal Timebase Mode: Lock to INT. 10 MHz
Trigger: External Direct Input, Level 50%
Acquisition: Average 64, Stop After Condition: Average Complete
Measurement: R1 to C1 Delay (R1/C1 Reference Level: Absolute
0.0 V)
2.
Click the
Inter-Channel Skew...
button in the Service UI menu to display the
following setup window:
Figure 3-7: Inter-Channel Skew Calibration setup window
Содержание AWG7000 Series
Страница 4: ......
Страница 10: ...List of Tables vi AWG7000 Series Service Manual...
Страница 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Страница 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Страница 18: ...Preface xiv AWG7000 Series Service Manual...
Страница 19: ...Operating Information...
Страница 20: ......
Страница 23: ...Theory of Operation...
Страница 24: ......
Страница 31: ...Adjustment Procedures...
Страница 32: ......
Страница 51: ...Maintenance...
Страница 52: ......
Страница 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Страница 109: ...Replaceable Parts...
Страница 110: ......
Страница 125: ...Replaceable Parts AWG7000 Series Service Manual 5 15 Figure 5 7 Exploded diagram Power supply assembly 1 2 3 4 5 6 6...
Страница 137: ...Replaceable Parts AWG7000 Series Service Manual 5 27 Figure 5 14 Exploded diagram SMA bracket 1 2 3 5 4 6 8 7...
Страница 139: ...Replaceable Parts AWG7000 Series Service Manual 5 29 Figure 5 15 Exploded diagram Relay unit 11 2 7 4 5 6 3 8 10 9 1 4...