Adjustment Procedures
3-2
AWG7000 Series Service Manual
Required Equipment
The following equipment, or a suitable equivalent, is required to complete these
procedures.
Adjustment Overview
The adjustment procedure contains up to ten adjustment items, depending on the
instrument type and options. Before performing adjustment procedures, you must
warm up the arbitrary waveform generator at least 20 minutes in an ambient
temperature between 20
°
C and 30
°
C.
Enable the Service Menu
You must enable the Service menu to perform the adjustment procedure. To do
this:
1.
Power on the instrument.
2.
Select the
System
menu from the menu bar, and then select
Service Mode
...
3.
The following dialog box is displayed. Enter the password “1185”.
Table 3-1: Test equipment
Description
Minimum requirements
Examples
Quantity
Sampling oscilloscope
Bandwidth: 20 GHz or higher
Tektronix CSA8200 with 80E03
1 ea
Oscilloscope
Bandwidth: 1 GHz or higher
Tektronix TDS5104B
1 ea
Frequency counter
Accuracy: within ± 0.01 ppm
Agilent 53181A
1 ea
50
Ω
BNC cable
DC to 2 GHz
Tektronix 012-0482-00
1 ea
50
Ω
SMA cable
DC to 20 GHz
Tensolite 1-3636-465-5236
2 ea
50
Ω
SMA Termination
DC to 18 GHz
Tektronix 015-1022-01
3 ea
SMA-BNC adapter
SMA male to BNC female
Tektronix 015-0554-00
2 ea
50
Ω
SMA Attenuator
DC to 18 GHz, 12 dB
Hirose AT-112
2 ea
Содержание AWG7000 Series
Страница 4: ......
Страница 10: ...List of Tables vi AWG7000 Series Service Manual...
Страница 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Страница 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Страница 18: ...Preface xiv AWG7000 Series Service Manual...
Страница 19: ...Operating Information...
Страница 20: ......
Страница 23: ...Theory of Operation...
Страница 24: ......
Страница 31: ...Adjustment Procedures...
Страница 32: ......
Страница 51: ...Maintenance...
Страница 52: ......
Страница 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Страница 109: ...Replaceable Parts...
Страница 110: ......
Страница 125: ...Replaceable Parts AWG7000 Series Service Manual 5 15 Figure 5 7 Exploded diagram Power supply assembly 1 2 3 4 5 6 6...
Страница 137: ...Replaceable Parts AWG7000 Series Service Manual 5 27 Figure 5 14 Exploded diagram SMA bracket 1 2 3 5 4 6 8 7...
Страница 139: ...Replaceable Parts AWG7000 Series Service Manual 5 29 Figure 5 15 Exploded diagram Relay unit 11 2 7 4 5 6 3 8 10 9 1 4...