Adjustment Procedures
3-8
AWG7000 Series Service Manual
b.
Modify the oscilloscope settings as follows:
CH 1 and CH 2 Vertical Scale: 200 mV/div
CH 1 and CH 2 Impedance: 50
Ω
.
Horizontal Scale: 20 ns/div
Trigger: Edge Trigger, CH 1/CH 2 Rising, Set Level to 50%
2.
Click the
Data Timing...
button in the Service UI menu to display the
following setup window:
Figure 3-5: Data Timing Calibration setup window
3.
Select
Step 4
in the Active Step field.
4.
Select
Ch 1
in the Channel field.