Adjustment Procedures
3-10
AWG7000 Series Service Manual
Regard the first or last settings where the waveform is broken as the end
point of broken waveform. In other words, if you have already click
Start
of Broken Waveform
, click
End of Broken Waveform
next.
If the waveform is broken with one set value only, and not across a range
of multiple settings, click both
Start of Broken Waveform
and
End of
Broken Waveform
at the same setting. Always click
Start of Broken
Waveform
first.
9.
Select
Ch 2
in the Channel field.
10.
Click
Set to 0
. This will reset the value of
TEK0015 Data Delay
to 0 ps.
11.
12.
Click
Update Cal Values
.
13.
Select
Ch 1
in the Channel field.
14.
Set
Frequency
to 10 GHz, and observe the oscilloscope screen. If a sine
waveform is not output normally, record the range as described in step 6 and
step 7.
15.
Decrease
Frequency
in the
Parameters for Step 0,4
field every 50 MHz from
10.000 GHz to 7.000 GHz using the up/down button or the front-panel rotary
knob. You can also enter the value directly using the keyboard.
16.
If the waveform starts to break at some point, click
Start of Broken
Waveform
. Click
End of Broken Waveform
right before the next waveform
is output normally (the last position where the waveform is broken in that
range).
17.
Continue the measurement to 7 GHz.
18.
Select
Ch 2
in the Channel field.
19.
Reset the value of
Frequency
to 10 GHz.
20.
Repeat step 14 through step 16.
21.
Click
Update Cal Values
.
22.
Click
Save to EEPROM
.
23.
Click
Close
. The Data Timing Calibration window closes.
Содержание AWG7000 Series
Страница 4: ......
Страница 10: ...List of Tables vi AWG7000 Series Service Manual...
Страница 14: ...Service Safety Summary x AWG7000 Series Service Manual...
Страница 16: ...Environmental Considerations xii AWG7000 Series Service Manual...
Страница 18: ...Preface xiv AWG7000 Series Service Manual...
Страница 19: ...Operating Information...
Страница 20: ......
Страница 23: ...Theory of Operation...
Страница 24: ......
Страница 31: ...Adjustment Procedures...
Страница 32: ......
Страница 51: ...Maintenance...
Страница 52: ......
Страница 58: ...Maintenance 4 6 AWG7000 Series Service Manual...
Страница 109: ...Replaceable Parts...
Страница 110: ......
Страница 125: ...Replaceable Parts AWG7000 Series Service Manual 5 15 Figure 5 7 Exploded diagram Power supply assembly 1 2 3 4 5 6 6...
Страница 137: ...Replaceable Parts AWG7000 Series Service Manual 5 27 Figure 5 14 Exploded diagram SMA bracket 1 2 3 5 4 6 8 7...
Страница 139: ...Replaceable Parts AWG7000 Series Service Manual 5 29 Figure 5 15 Exploded diagram Relay unit 11 2 7 4 5 6 3 8 10 9 1 4...