:SENSe Commands
3- 28
AVG1 Analog Video Generator User Manual
:SENSe:CORRection:MDELay:HORizontal(?)
Use this command to set or query the horizontal timing delay of the module with
respect to the rest of the instrument.
Argument units are microseconds. If the argument exceeds the horizontal time
value, then the vertical delay is adjusted to accommodate the requested delay.
Settings are saved through power cycles.
:SENSe:CORRection:MDELay:HORizontal <numeric_value>
:SENSe:CORRection:MDELay:HORizontal?
Command
Query response
<numeric_value> = <NRf>
UP, DOWN, DEF
<NR2>
0.0
--200 Execution error
None
Command:
:SENS:CORR:MDEL:HOR 23.0
Query:
:SENS:CORR:MDEL:HOR?
Response:
23.0000
:SENSe:CORRection:MDELay:HORizontal
Syntax
Parameters
Default Value
Errors and Events
Dependencies
Examples
Related Commands
Содержание AVG1
Страница 4: ......
Страница 10: ...Table of Contents vi AVG1 Analog Video Generator User Manual...
Страница 14: ...Service Safety Summary x AVG1 Analog Video Generator User Manual...
Страница 17: ...Getting Started...
Страница 18: ......
Страница 23: ...Operating Basics...
Страница 24: ......
Страница 42: ...Operating Procedures 2 18 AVG1 Analog Video Generator User Manual...
Страница 43: ...Syntax and Commands...
Страница 44: ......
Страница 48: ...Syntax 3 4 AVG1 Analog Video Generator User Manual...
Страница 70: ...OUTPut Commands 3 26 AVG1 Analog Video Generator User Manual...
Страница 76: ...SENSe Commands 3 32 AVG1 Analog Video Generator User Manual...
Страница 115: ...Appendices...
Страница 116: ......
Страница 124: ...Appendix B SCPI Conformance Information B 4 AVG1 Analog Video Generator User Manual...
Страница 126: ......
Страница 138: ...Appendix C Installation C 12 AVG1 Analog Video Generator User Manual...
Страница 139: ...Glossary and Index...
Страница 140: ......
Страница 144: ...Glossary Glossary 4 AVG1 Analog Video Generator User Manual...