AFG1000 Series Arbitrary Function Generator Programmer Manual
21
Command Descriptions
[SOURce[1|2]]:AM[:DEPTh]
This command sets or queries the modulation depth of AM modulation for the
specified channel. Set the modulation depth from 0% to 100% with resolution of
1%.
Conditions
For the AFG1022, use
[SOURce1]
. Modulation, sweep, and burst conditions do
not apply to CH2 for the AFG1022.
Group
Source
Syntax
[SOURce[1|2]]:AM[:DEPTh]
{<depth>|MINimum|MAXimum}
[SOURce[1|2]]:AM[:DEPTh]
? [MINimum|MAXimum]
Arguments
<depth>::=<NR2>[<units>]
where:
<NR2> is the depth of modulating frequency.
<units>::=PCT
MINimum sets the modulation depth to minimum value.
MAXimum sets the modulation depth to maximum value.
Returns
<depth>
Examples
SOURce1:AM:DEPth MAXimum
sets the depth of modulating signal on CH 1 to the maximum value.
[SOURce[1|2]]:AM:INTernal:FREQuency
This command sets or queries the internal modulation frequency of AM
modulation for the specified channel. Use this command when the internal
modulation source is selected. Set the internal modulation frequency from 2
mHz to 20.00 kHz with resolution of 1 mHz.
Conditions
For the AFG1022, use
[SOURce1]
. Modulation, sweep, and burst conditions do
not apply to CH2 for the AFG1022.