Command Descriptions
AFG1000 Series Arbitrary Function Generator Programmer Manual
Returns
<OPT>[,<OPT>[,<OPT>[,<OPT>]]]
Examples
*OPT?
might return 0, which indicates no option is installed in the instrument.
OUTPut[1|2]:IMPedance
This command sets the output load impedance for the specified channel. The
specified value is used for amplitude, offset, and high/low level settings. You can
set the impedance to any value from 1 Ω to 10 kΩ with a resolution of 1 Ω. The
default value is 50 Ω.
The query returns the current load impedance setting in ohms. If the load
impedance is set to INFinity, the query returns “9.9E+37”.
Group
Output
Syntax
OUTPut[1|2]:IMPedance
{<ohms>|INFinity|MINimum|MAXimum}
OUTPut[1|2]:IMPedance?
{MINimum|MAXimum}
Arguments
<ohms>::=<NR3>[<units>]
where:
<units>::=OHM
INFinity sets the load impedance to >10 kΩ.
MINimum sets the load impedance to 1 Ω.
MAXimum sets the load impedance to 10 kΩ.
Returns
<ohms>::=<NR3>
Examples
OUTPut1:IMPedance MAXimum
sets the CH 1 load impedance to 10 kΩ.
OUTPut[1|2][:STATe]
This command sets or query the instrument output state for the specified
channel.