![Tektronix 2245A Скачать руководство пользователя страница 76](http://html.mh-extra.com/html/tektronix/2245a/2245a_operators-manual_1078849076.webp)
Basic Applications
FREQUENCY
Figure 5-3. Frequency measurement.
R is e -T im e M e a su re m en ts
Making rise-time or fail-time measurements requires some additional signal
scaling to use the graticule rise-time measurement aids. Numbers 0%, 10,
90, and 100 are etched near the left vertical graticule line. These numbers
provide convenient reference points for making rise- and fall-time measure
ments when the signal to be measured is properly set up. Use the following
steps as a guideline In making rise-time measurements.
1. Apply the signal to the CH 1 input connector.
2. Set:
Vertical MODE
CH 1
Horizontal MODE
A
A Trigger SLOPE
_ r
5-6