![Tektronix 2245A Скачать руководство пользователя страница 128](http://html.mh-extra.com/html/tektronix/2245a/2245a_operators-manual_1078849128.webp)
Performance Check Procedure
12. Vertical POSITION Range (all channels)
Set:
A SEC/DIV
VERTICAL MODE
CH 1 VOLTS/DIV
CH 2 INVERT
SCOPE BW
0.1 ms
CH 1 (ADD off)
1 V
Off
Off
b.
c.
d.
e.
f.
g-
h.
i.
j.
k.
l
.
CH 1 and CH 2 Input COUPLING
AC
Connect leveled sine-wave generator (SG 503) output to the CH 1 and
CH 2 inputs via a 50
f l
BNC coaxial cable, a 50 H BNC termination,
and a BNC dual-input coupler.
Position trace to center horizontal graticule line.
Set leveled sine-wave generator output for two-division signal at
50 kHz.
Set:
CH 1 VOLTS/DIV
0.1 V
CH 1 POSITION
Fully CW (clockwise)
CHECK—that the bottom of the waveform is at least one division above
the center horizontal graticule line.
Set CH 1 POSITION fully CCW.
CHECK—that the top of the waveform is at least one division below the
center horizontal graticule line.
Set:
CH 1 POSITION
12 o ’ clock
VERTICAL MODE
CH 2 (CH 1 off)
CH 2 POSITION
Fully CW
CHECK—that the bottom of the waveform is at least one division above
the center horizontal graticule line.
Set CH 2 POSITION fully CCW.
CHECK—that the top of the waveform is at least one division below the
center horizontal graticule line.