P7350 5 GHz Differential Probe Instruction Manual
29
Appendix B: Performance Verification
Use the following procedures to verify specifications of the probe.
Before beginning these procedures, refer to page 42 and photocopy
the test record, and use it to record the performance test results. The
recommended calibration interval is one year.
These procedures test the following specifications:
H
Output offset voltage
H
DC gain accuracy
H
Rise time
Equipment Required
Refer to Table 7 for a list of the equipment required to verify the
performance of your probe.
Table 7: Equipment required for performance verification
Item description
Performance requirement
Recommended example
1
Sampling Oscilloscope
20 GHz bandwidth
Tektronix TDS8000
Sampling module (head),
with extension cable
20 GHz bandwidth
Tektronix 80E04 with
012-1568-00 cable
Oscilloscope
TekConnect interface
Tektronix TDS7404
Calibration Step Generator
067-1338-0X
Probe Positioner
Tektronix PPM203B
Dual Power Supply
5.0 VDC at 1 mA
Tektronix PS280
DMM (2), with leads
0.1 mV resolution
Fluke 87 or equivalent
Feedthrough Termination
BNC, 50
Ω
±
0.05
Ω
011-0129-00
Coaxial cable
Male-to-Male SMA
012-0649-00