OPERATING INSTRUCTIONS
The resistance reading will show up on the display. If the
measurement is over-range, the display will read “1" followed
by three blanks. In that case, turn the range switch to the next
higher range.
When the battery voltage has fallen to a value near the end of
its useful life, “LOBAT” will be seen on the display. The unit
will not be damaged by further use, but this is a warning that
readings may not be accurate, especially on the 2 mΩ range
which uses a nominal 1.3 A test current. The battery cells
should be replaced when “LOBAT” appears.
Depress the POWER ON/OFF switch to OFF, when through
using the instrument, to avoid depleting the battery.
Disconnect the test probes and store them in the case lid with
the instruction manual, upon completion of the measurement
task.