Page 2-2
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • [email protected]
PREPARATION FOR USE AND INSTALLATION
R1L-E2A are listed on the Type Certificate and are:
o
Energizer E95
2.2.2 Test Probes
Only HTP-101A or SKP-8 test probes or leads may be used in
an explosive atmosphere.
2.3 Preparation for Use
Note that the Model R1L-E2A is shipped with its battery cells
removed. Three battery cells are shipped with each new unit.
•
Loosen the six captive screws around the outside edge of
the battery access panel using a Phillips No. 2 screwdriver,
and remove the panel.
•
Check for correct polarity and install the three D-cells in the
battery holder. When installing the cells, rotate them slightly
to ensure connection to the battery holder contacts. The
cells are connected in series for a nominal voltage of 4.5 V.
WARNING
DO NOT INSTALL, CHANGE, OR REMOVE THE R1L-E2A
BATTERY CELLS IN AN EXPLOSIVE ATMOSPHERE! A SPARK
MIGHT BE GENERATED, WHICH COULD RESULT IN AN
EXPLOSION.
WARNING
BATTERIES ARE PART OF THE INTRINSICALLY SAFE
SYSTEM. USE ONLY RECOMMENDED BATTERIES. DO NOT
MIX BATTERIES. USE OF ANY OTHER BATTERY OR MIXED
BATTERIES MAY RESULT IN A HAZARDOUS CONDITION.
WARNING
USE ONLY HTP-101A OR SKP-8 TEST PROBES AND CABLE
SETS WITH R1L-E2A IN POTENTIALLY HAZARDOUS
ENVIRONMENTS. USE OF ANY OTHER COMBINATION OF
PROBE AND CABLE INVALIDATES THE TYPE CERTIFICATION
AND MAY RESULT IN A HAZARDOUS CONDITION.