10 TEGAM WAY • GENEVA, OHIO 44041 440-466-6100 • FAX 440-466-6110 • www.tegam.com
1-5
Section I – Introduction
Items Included with the AVM-2000
Item Part
Number
AC Power Module
Rechargeable Battery
Input Terminal Shield
GUARD to LO Nickel Plated Shorting Link
2
μ
F Filter Block Mounted on Dual Banana Jack
Technical (Instruction and Service) Manual
PS-107
BA-105
0600-081
J-5
0600-090
AVM2000-900
Table 1.2— AVM-2000 Packing List
Additional items required for operation and maintenance
Precision voltage and current source 0 to 1000 Volts DC & AC RMS and 10
μ
A – 1 A DC, such as
Fluke model 5100 or 5720 Multi Product Calibrator
Precision DMM with range of 0 to ±1000 V DC and AC RMS such as Fluke model 8508A, Keithley
2002 or Agilent 3458A.
Oscilloscope such as Tektronix model TDS-1000
Environmental chamber capable of housing the AVM-2000 with ability to regulate temperature from
10°C to 35°C ±0.1°C such as Thermotron model S-5.5 3800
Megohmmeter such as PPM model R1M-A
Low-thermal EMF shielded test leads (solid copper wire or solid copper wire with gold plated spade
terminals)
Pure copper shorting strap
10 m
Ω
10 A shunt, 0.1%
25
Ω
resistor, ¼ Watt, 1%
30
Ω
resistor, ¼ Watt, 1%
10 k
Ω
resistor. ¼ Watt, 1%
1 M
Ω
resistor, ¼ Watt, 1%
10 M
Ω
resistor, ¼ Watt, 2%
100 M
Ω