Section V – Maintenance and Servicing
Calibration Verification (Preliminary Calibration Operations)
1.
Connect test equipment to appropriate power sources, set POWER switches to ON and allow a
warm-up period (Suggest 4 hours).
NOTE: Recommend use of low-thermal EMF shielded test leads (solid copper wire or solid copper wire with
gold plated spade terminals).
NOTE: Before continuing, ensure AVM-2000 has been connected to external power for a minimum of 20
hours.
2.
With AVM-2000 POWER switch set to OFF, disconnect AVM-2000 from external power source and
mechanically zero AVM-2000 meter.
NOTE:
The AVM-2000 must remain at the same tilt level during warm-up and throughout the Calibration
Process.
3.
Set AVM-2000 POWER switch to On, ensure AVM-2000 indicates a fully charged battery and allow a
one hour warm-up period.
4.
Connect the Digital Multimeter HI and LO INPUT terminals to AVM-2000 rear panel + and -
terminals, observe polarity.
5.
Set the Digital Multimeter FUNCTION for DCV and AUTO range.
If AVM-2000 LCD OFST: indication is not 00,000.00
μ
V, using the INPUT OFFSET knob to select the
place value and the OFFSET
▼
▲
buttons to select the numerical value, set OFST: indication to
00,000.00
μ
V.
6.
Press AVM-2000 INPUT OFFSET knob until V is displayed. Press and hold AVM-2000 INPUT OFFSET
knob for approximately 5 seconds.
7.
Press AVM-2000 INPUT OFFSET knob until I is displayed. Press and hold AVM-2000 INPUT OFFSET
knob for approximately 5 seconds.
8.
Press AVM-2000 INPUT OFFSET knob until a padlock symbol is displayed in the lower right of the
LCD.
9.
If required, press AVM-2000 OPERATE/ZERO button to select IMPEDANCE: ZERO as indicated on
AVM-2000 LCD.
10.
Press and hold AVM-2000 RANGE knob for approximately 5 seconds until AVM-2000 LCD indicates
METER OFFSET.
11.
Using AVM-2000 OUTPUT LEVEL knob, adjust AVM-2000 meter for a zero indication.
12.
Turn AVM-2000 RANGE knob one step clockwise and AVM-2000 LCD readout will indicate METER
GAIN +.