Section V – Maintenance and Servicing
Range (VDC) Applied (VDC)
Accuracy Limits (VDC)
1
μ
±1
μ
±0.980 to 1.020
μ
3
μ
±3
μ
±2.940 to 3.060
μ
10
μ
±10
μ
±9.80 to 10.20
μ
30
μ
±30
μ
±29.40 to 30.60
μ
100
μ
±100
μ
±98.0 to 102.0
μ
300
μ
±300
μ
±294.0 to 306.0
μ
1 m
±1 m
±0.980 to 1.020 m
3 m
±3 m
±2.940 to 3.060 m
10 m
±10 m
±9.80 to 10.20 m
30 m
±30 m
±29.40 to 30.60 m
100 m
±100 m
±98.0 to 102.0 m
300 m
±300 m
±294.0 to 306.0 m
1
±1
±0.980 to 1.020
3
±3
±2.940 to 3.060
10
±10
±9.80 to 10.20
30
±30
±29.40 to 30.60
100
±100
±98.0 to 102.0
300
±300
±294.0 to 306.0
1000
±1000
±980.0 to 1020.0
1
0
Ref
±1
Ref
±0.8
±2%
±0.6
±2%
±0.4
±2%
±0.2
±2%
3
0
Ref
±3
Ref
±2
±2%
±1
±2%
Table 5.3 Analog Meter Calibration Performance Table
Range (VDC)
Applied (mV DC)
Linearity Limits
1 mV
0
Ref
±1
Ref
±0.8
±0.5%
±0.6
±0.5%
±0.4
±0.5%
±0.2
±0.5%
Table 5.4 Isolated Output Calibration Performance Table
Range (VDC)
Applied (VDC) Accuracy Limits (VDC)
30 m
±27 m
±27.15000 to 26.85000 m
Table 5.5 Offset Voltage Calibration Performance Table