Service Information
5
-
5
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440
-466-
6100 ● FAX 440
-466-
6110 ● [email protected]
25.
The UUT display will indicate as follows:
a.
“1”
b.
Line 1:
-----
c.
Line 2:
1
Ω
26.
Short the leads of the unit as described in
Step 6.
27.
Press the
to select 0.000
28.
Press
a.
“-----
” will flash periodically
.
29.
The UUT display will then indicate as follows:
a.
“2”
b.
Line 1:
-----
c.
Line 2:
1
Ω
30.
Connect the probes to the
1
Ω
resistance standard
.
31.
A value will be displayed, press the
to select the correct value of the 1 Ω
standard.
32.
Press
a.
“-----
” will flash periodically
33.
The UUT display will indicate as follows:
a.
“1”
b.
Line 1:
-----
c.
Line 2: 10
Ω
34.
Short the leads of the unit as described in
Step 6
.
35.
Press
to select 0.000.
36.
Press
a.
“-----
” will flash periodically
37.
The UUT display will then indicate as follows:
a.
“2”
b.
Line 1:
-----
c.
Line 2: 10
Ω
38.
Connect the probes to the 10
Ω
resistance standard