background image

Operating manual 
E3907 
 

www.tecsis.de 

BE 984 i 

13 

 
 

4.6.2.2 Read-Out of current sensor parameters: 
 
Sensor designation 

8Byte 

ASCII 

Display final value 

2Byte 

packed BCD-figure 

Unit 

3Byte 

ASCII 

Sensor type and digit 

1Byte 

0xAB:  A ... Sensor type, B ... digit of the dec. point (binary  coded) 
Sensor type: 
 0xXXXX XXXX 
  |||| 
  0000 ... active   with calibration control 0%load and 100% load calibration 
  0001 ... active   without calibration control 0%load and 100% load calibration 
  0010 ... active   without calibration control 0%load calibration and edit 100%load V 
  0011 ... active   without calibration control edit 0%load V and 100%load V  
  0100 ... passive  with calibration control 0%load and 100% load calibration 
  0101 ... passive  without calibration control 0%load and 100% load calibration 
  0110 ... passive  without calibration control 0%load calibration and edit 100% load in mV/V 
  0111 ... passive  without calibration control edit 0%load and 100% load in mV/V 
  1000 ... current                 0%load and 100% load calibration 
  1001 ... current                 0%load calibration and edit 100%load in mA 
  1010 ... current                 edit 0%load mA and 100%load mA  
 
Dec. point: 
0xXXXX XXXX 
       |||| 
       |000 ... _5000___ 
       |001 ... _5,000__ 
       |010 ... _50,00__ 
       |011 ... _500,0__ 
       |100 ... 5,000___ 

0%load 

2Byte 

HEX-value (MSB/LSB) 

100% load 

2Byte 

HEX- value (MSB/LSB) 

 
 

4.6.2.3 Read out status: 
Status  2Byte, general error condition of the E3907 
 
4.6.2.4 Read-Out complete status: 
Status 

2Byte 

general error condition of the GM 80 

Measuring rate 

1Byte 

0x01 … 1000/sec 
0x02 … 100/sec 
0x03 ...  10/sec 
0x04 ...   1/sec 

Average value 

1Byte 

0x01 ... x/1 
0x02 ... x/2 
0x04 ... x/4 
0x08 ... x/8 
0x10 ... x/16 
0x20 ... x/32 

SCI_MODE 

1Byte 

0x00 ... interface off 
0x04 … hand mode 
0x08 … automatic mode 
0x0C … trigger mode 

SCI_MODE_DELAY 

1Byte 

0x02 …10ms 
0x03 … 100ms 
0x04 … 1s 
0x05 … 10s 
0x06 ... 1min 
0x07 ... 10min 
0x08 ... 1h 

LOGGMODE 

1Byte 

0x00 ... logger off 
0x04 ... hand mode 
0x08 ... automatic mode 
0x0C ... graph mode 
0x10 ... screen mode 

LOGGMODE_DELAY 

1Byte 

0x01 … 1ms 
0x02 …10ms 
0x03 … 100ms 
0x04 … 1s 
0x05 … 10s 
0x06 ... 1min 
0x07 ... 10min 
0x08 ... 1h 

Language 

1Byte 

0x00 ... GERMAN 
0x02 ... ENGLISH 
0x04 ... FRENCH 
0x06 ... SPANISH 

Protocol status 

1Byte 

0xXXXX XXXX (binary coded) 
  |||| |||| 
  |||| |||1 ... do not send  conclusion character 
  |||| ||1  ... send CR/LF  
  |||| |1   ... send CR   
  |||| 1    ... send LF

  

 

 

 
4.6.2.5 Read time 
The writing is identically with the data block for receipt of time, however, the data block for writing is protected with a 
checksum and the corresponding weighted checksum  
 
 
 

Содержание E3907

Страница 1: ...Force Pressure Temperature Switch Service Operating manual E3907 Universal handheld measuring device with datalogger BD 984 i 02 2013...

Страница 2: ...y 63073 Offenbach am Main Valid for E3907 handheld measuring device with datalogger Copyright 2006 2015 tecsis GmbH Offenbach Germany Reproduction prohibited Reproduction in whole or in part only with...

Страница 3: ...interface 11 4 6 1 Transfer rate specification RS232 V 24 USB 11 4 6 2 Operation via the serial interface 11 4 6 3 Handling with serial interface USB 14 4 6 4 Operation and function principle of the E...

Страница 4: ...Operating manual E3907 BE 984 i www tecsis de...

Страница 5: ...arantee Errors and changes are reserved Intended Purpose Improper Usage A tecsis device is used for displaying processing and controlling or regulation of processes It shall not be used as the only to...

Страница 6: ...take responsibility for that residual risks and to keep them as low as possible The relevant regulations and laws must be observed Residual risks or dangers associated with the measurement device shou...

Страница 7: ...s The device may only be used by qualified personnel according to the technical data in connection with the special safety rules and regulations 1 8 Technical changes The product may not be changed co...

Страница 8: ...An assignment of sensors to the corresponding parameter set is documented on the rear of the device 3 2 Power supply Mains operation Power supply 6V DC min 1 8A Fuse Self reset able fuse 1 5A Under Ov...

Страница 9: ...evice from moisture condensation rain snow Protect the device from direct sunlight Protect the device from dust and dirt Protect the device from high or excessive ambient temperature Protect the devic...

Страница 10: ...nu ESC Switch off the device Exit one step back in the menu Scroll up Scroll down USB USB Socket for USB Connector Type B RS232 RS232 Socket for 3 5mm jack 3pol front GND middle RxD outside TxD 6V DC...

Страница 11: ...Operating manual E3907 www tecsis de BE 984 i 7...

Страница 12: ...Operating manual E3907 8 BE 984 i www tecsis de 4 2 Switching on the device Press the ENTER button to switch on the device 4 3 Menu description 4 3 1 Brief handling overview E3907...

Страница 13: ...w many measurements an average value is accumulated Applicable at e g vibrations control oscillations 2 5 PASS Password query After entering the password 9373 it is possible to enter menu 3 CALIB Ther...

Страница 14: ...appear on the LCD Press ENTER again and the time can be adjusted With PLUS and MINUS the hours can be adjusted After ENTER the minutes can be adjusted with PLUS and MINUS With ENTER the selected time...

Страница 15: ...uring every keystroke on ENTER while measuring For data logging adjust operating mode in the logger mode e g choose AUTO and interval time e g 10 sec and change to measuring mode A measured value with...

Страница 16: ...Out datalogger B 0x42 Delete datalogger C 0x43 Read Out current sensor parameter D 0x44 Read Out status E 0x45 Read Out complete status g 0x67 Change protocol setup k 0x6B ENTER l 0x6C PLUS m 0x6D MI...

Страница 17: ...mA and 100 load mA Dec point 0xXXXX XXXX 000 _5000___ 001 _5 000__ 010 _50 00__ 011 _500 0__ 100 5 000___ 0 load 2Byte HEX value MSB LSB 100 load 2Byte HEX value MSB LSB 4 6 2 3 Read out status Status...

Страница 18: ...ol status 1Byte 0xXXXX XXXX binary coded 1 no closing character is sent 1 send CR LF 1 send CR 1 send LF 4 6 2 12 Calculation of the Checksum CS and the Weighted Checksum wCS The calculation takes pla...

Страница 19: ...starting time of the measurement the current sensor designation the final value of the measuring range the adjusted measuring rate and the logger mode e g AUTO 1ms are saved in the logger NOTE All pr...

Страница 20: ...ress key or and choose between 0 Load or Sensitivity 0 Load Relieve sensor from load or Sensitivity Enter the value fort the sensitivity in V Press Key or and choose between 100 Load or Sensitivity 10...

Страница 21: ...0mA short circuit protected Bridge resistance min 350 Ohm only full bride 4 or 6 conductor Sensitivity up to 3 3mV V other values on request 100 control via E3907 applied via control input 4 8 2 Conne...

Страница 22: ...measures before installation Attention The information for the voltage supply in section 5 2 must be respected Cable connections Attention Never connect voltage supply on non used pins 5 3 Normal ope...

Страница 23: ...k the plug connections Repair Notation The products contain no parts that need to be or that can be serviced by the user himself Repairs may only be carried out by the tecsis GmbH If it can be assumed...

Страница 24: ...tecsis GmbH Carl Legien Stra e 40 44 63073 Offenbach am Main Germany Phone 49 69 5806 0 Fax 49 69 5806 7788 E Mail force tecsis de Internet www tecsis de...

Отзывы: