If the
Alarm System
is
ON
, when alarm occurs, the oximeter gives visual alarm and audio
alarm.
Press
Alarm Silence
button can pause the alarm system for seconds (the paused period
can be set to 30, 60, 90 or 120s by user, see 3.3.5), the Audio alarm off icon displays. But
the visual alarm is still active. For example, if the measured SpO
2
value is higher than
SpO
2
Hi Alarm
or lower than
SpO
2
Lo Alarm
, there will be
↑
or
↓
icon display on screen, and
the SpO
2
or PR character will flash.
If the
Alarm system
is set to
OFF
, all the audio alarm and visual alarm are turned off.
WARNING
When the Alarm system is off, the oximeter will not give alarm prompt. In order to avoid
endangering the patient’s life, the user should use this function cautiously.
3.3.9 SpO
2
Alarm setup
User can choose
SpO
2
Hi Alarm
and
SpO
2
Lo Alarm
in menu to adjust SpO
2
alarm limit.
Press
Up Arrow
button or
Down Arrow
button to increase or decrease alarm limit.
In default configuration setup, the
SpO
2
Hi Alarm
and
SpO
2
Lo Alarm
for
Neo
mode are
95 and 90; while they are 100 and 90 for
Adu
mode.
Set the SpO
2
alarm limits as following procedure:
Choose
SpO
2
Hi Alarm
in the menu, press
Function
button to enter setup. The
SpO
2
Hi Alarm
box will change from real line box to broken line box. The
adjustable range for upper limit of SpO
2
is from 1 plus the lower limit of SpO
2
to
100.
Press
Up Arrow
or
Down Arrow
button to increase or decrease values.
Choose
SpO
2
Lo Alarm
in the menu, press
Function
button to set up. The
SpO
2
Lo Alarm
box will change from real line box to broken line box. The adjustable